Sensors (Basel). 2023 May 17;23(10):4820. doi: 10.3390/s23104820.
Force plate testing is becoming more commonplace in sport due to the advent of commercially available, portable, and affordable force plate systems (i.e., hardware and software). Following the validation of the Hawkin Dynamics Inc. (HD) proprietary software in recent literature, the aim of this study was to determine the concurrent validity of the HD wireless dual force plate hardware for assessing vertical jumps. During a single testing session, the HD force plates were placed directly atop two adjacent Advanced Mechanical Technology Inc. in-ground force plates (the "gold standard") to simultaneously collect vertical ground reaction forces produced by 20 participants (27 ± 6 years, 85 ± 14 kg, 176.5 ± 9.23 cm) during the countermovement jump (CMJ) and drop jump (DJ) tests (1000 Hz). Agreement between force plate systems was determined via ordinary least products regression using bootstrapped 95% confidence intervals. No bias was present between the two force plate systems for any of the CMJ and DJ variables, except DJ peak braking force (proportional bias) and DJ peak braking power (fixed and proportional bias). The HD system may be considered a valid alternative to the industry gold standard for assessing vertical jumps because fixed or proportional bias was identified for none of the CMJ variables (n = 17) and only 2 out of 18 DJ variables.
由于商用、便携式和价格合理的力量板系统(即硬件和软件)的出现,力量板测试在运动中变得越来越普遍。在最近的文献中验证了 Hawkin Dynamics Inc.(HD)专有的软件之后,本研究的目的是确定 HD 无线双力量板硬件评估垂直跳跃的同时效度。在一次单独的测试过程中,HD 力量板直接放在两个相邻的 Advanced Mechanical Technology Inc. 地下力量板(“黄金标准”)上,同时收集 20 名参与者(27±6 岁,85±14 公斤,176.5±9.23 厘米)在反跳(CMJ)和跳落(DJ)测试(1000 Hz)期间产生的垂直地面反作用力。使用Bootstrapped 95%置信区间的普通最小二乘回归确定了两个力量板系统之间的一致性。在任何 CMJ 和 DJ 变量中,两个力量板系统之间都没有出现偏差,除了 DJ 峰值制动力(比例偏差)和 DJ 峰值制动力(固定和比例偏差)。HD 系统可被视为评估垂直跳跃的行业黄金标准的有效替代方案,因为在 17 个 CMJ 变量中没有一个(n=17),在 18 个 DJ 变量中只有 2 个出现固定或比例偏差。