评价两种列线图和四种预测公式在可植入 Collamer 透镜尺寸选择中的性能。

Evaluation of the Performance of Two Nomograms and Four Vault Prediction Formulas for Implantable Collamer Lens Size Selection.

出版信息

J Refract Surg. 2023 Jul;39(7):456-461. doi: 10.3928/1081597X-20230605-01. Epub 2023 Jul 1.

Abstract

PURPOSE

To evaluate the performance of different nomograms and vault prediction formulas in predicting the optimal Implantable Collamer Lens (ICL; STAAR Surgical) size and vault.

METHODS

This retrospective study included 108 participants (214 eyes) who underwent ICL implantation. The efficacy of the manufacturer's nomogram, Parkhurst nomogram, NK formula, KS formula, Zhu formula, and ZZ formula was assessed by comparing the indicated ICL sizes to the postoperative vault measurements. Additionally, Bland-Altman plots and the Friedman test were used to assess the agreement and absolute error between the actual vault and predicted vault.

RESULTS

The proportions of the ideal ICL category recommended by the manufacturer's nomogram, Parkhurst nomogram, NK formula, KS formula, Zhu formula, and ZZ formula were 50.5%, 45.3%, 46.7%, 42.5%, 50.0%, and 28.5%, respectively. The mean differences between the actual vault and the predicted vault using the NK, KS, Zhu, and ZZ formulas were 144.1 ± 261.1, -19.3 ± 179.6, 70.8 ± 284.2, and 182.6 ± 361.5 μm, respectively. The predicted ICL vault tended to overestimate the actual ICL vault, particularly when choosing a larger ICL size. The KS formula shows the smallest deviation in prediction error and is least affected by variation in ICL size.

CONCLUSIONS

White-to-white distance from the Pentacam (Oculus Optikgeräte GmbH) coupled with the manufacturer's nomogram performed well for ICL size selection. Four vault prediction formulas tended to overestimate the actual ICL vault, particularly when selecting a larger ICL size. The KS formula appeared to have the least bias of the formulas. Simultaneously, vault prediction formulas need to be modified according to the ICL size. .

摘要

目的

评估不同的列线图和 vault 预测公式在预测最佳可植入 Collamer 透镜(ICL;STAAR Surgical)大小和 vault 方面的性能。

方法

本回顾性研究纳入了 108 名(214 只眼)接受 ICL 植入的患者。通过比较制造商的列线图、Parkhurst 列线图、NK 公式、KS 公式、Zhu 公式和 ZZ 公式所预测的 ICL 大小与术后 vault 测量值,评估了这几种方法的有效性。此外,还使用 Bland-Altman 图和 Friedman 检验评估了实际 vault 和预测 vault 之间的一致性和绝对误差。

结果

制造商的列线图、Parkhurst 列线图、NK 公式、KS 公式、Zhu 公式和 ZZ 公式推荐的理想 ICL 类别比例分别为 50.5%、45.3%、46.7%、42.5%、50.0%和 28.5%。使用 NK、KS、Zhu 和 ZZ 公式时,实际 vault 和预测 vault 之间的平均差异分别为 144.1±261.1、-19.3±179.6、70.8±284.2 和 182.6±361.5μm。预测的 ICL vault 倾向于高估实际的 ICL vault,尤其是在选择较大的 ICL 尺寸时。KS 公式的预测误差偏差最小,受 ICL 尺寸变化的影响最小。

结论

Pentacam(Oculus Optikgeräte GmbH)的角膜顶点到角膜顶点距离结合制造商的列线图在 ICL 尺寸选择方面表现良好。四个 vault 预测公式倾向于高估实际的 ICL vault,尤其是在选择较大的 ICL 尺寸时。KS 公式似乎是这些公式中偏差最小的。同时,需要根据 ICL 尺寸修改 vault 预测公式。