Radiation Oncology, University of Washington, Seattle, Washington, USA.
Kaiser Permanente, Seattle, Washington, USA.
J Appl Clin Med Phys. 2023 Nov;24(11):e14094. doi: 10.1002/acm2.14094. Epub 2023 Jul 19.
Five complete MR-conditionally approved ring sets, including fifteen tandems, and two additional rings, were commissioned at an institution intending to use them in an MRI planning environment with a Bravos HDR brachytherapy remote afterloader. Channel length, radiograph, autoradiograph, ring offset, and treatment interrupt measurements were performed, and applicators were assessed in both CT and MRI. During commissioning, one ring was found to be defective and was returned to the manufacturer for a replacement. The eventual complete applicator suite (including the replacement ring) was found to follow the manufacturer-provided specifications, including those delineated in vendor-provided 3D virtual models and those defined within the manufacturer's instructions for use documentation. Based on this work, an offset correction of -0.4 cm will be used for all tested rings using the Bravos system's internal distal dwell position correction feature during treatment preparation. This study reiterated the requirement for careful commissioning of each applicator intended for clinical service considering the intended use and the planned clinical environment and work processes.
五个完整的 MR 条件批准的环套件,包括十五个串联件和两个额外的环,在一家机构中委托使用,该机构打算在具有 Bravo HDR 近距离放射治疗后装器的 MRI 计划环境中使用它们。进行了通道长度、射线照相、自动照相、环偏移和治疗中断测量,并在 CT 和 MRI 中评估了施源器。在委托过程中,发现一个环有缺陷,并返回制造商更换。最终完整的施源器套件(包括替换环)被发现符合制造商提供的规格,包括供应商提供的 3D 虚拟模型中规定的规格以及制造商使用说明文档中规定的规格。基于这项工作,将在治疗准备期间使用 Bravo 系统的内部远端驻留位置校正功能,对所有测试环使用-0.4cm 的偏移校正。这项研究重申了在考虑预期用途和计划的临床环境和工作流程的情况下,对每个计划用于临床服务的施源器进行仔细委托的要求。