比较口腔内扫描仪和面部扫描仪与经过验证的基准线在裂隙模型中对鼻唇区域的 3D 评估。

3D assessment of the nasolabial region in cleft models comparing an intraoral and a facial scanner to a validated baseline.

机构信息

Department of Oral and Cranio-Maxillofacial Surgery, Friedrich-Alexander-Universität Erlangen-Nürnberg, Glueckstrasse 11, 91054, Erlangen, Germany.

Department of Prosthodontics, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.

出版信息

Sci Rep. 2023 Jul 27;13(1):12216. doi: 10.1038/s41598-023-39352-7.

Abstract

We aimed to validate the metric accuracy of a 3-dimensional (3D) facial scanner (FS) and an intraoral scanner (IOS) in capturing the nasolabial region in ex vivo unilateral cleft lip and palate (UCLP) models. The nasolabial region of 10 UCLP models was scanned using a 3D FS as well as an IOS and a previously validated stationary 3D scanner as a reference. Intraoral scan was performed directly on the UCLP models. In order to apply the FS on the models, they were embedded in a 3D printed sample face. Both test groups were aligned to the reference by applying a section-based best-fit algorithm. Subsequent analysis of the metric deviation from the reference was performed with a 3D analysis tool. Mean distance and integrated distance served as main parameters for surface and volume comparison. Point comparison served as an additional parameter. Statistical analysis was carried out using t-test for unconnected samples. Considering mean distance and integrated distance as main parameters for 3D evaluation of the scanner's accuracy, FS and IOS differ significantly in their metric precision in scanning the cleft model compared to the reference. The IOS proved to be significantly more accurate than the FS compared to the previously described stationary 3D scanner as reference and validated baseline. Further validation of the tested IOS and FS for 3D assessment of the nasolabial region is presented by adding the previously validated ATOS III Triple Scan blue light scanner as a reference. The IOS shows, compared to a validated baseline scan, significantly higher metric precision in experimental cleft model scanning. The collected data provides a basis for clinical application of the IOS for 3D assessment of the nasolabial region.

摘要

我们旨在验证一种 3 维(3D)面部扫描仪(FS)和一种口内扫描仪(IOS)在获取单侧唇裂腭裂(UCLP)模型的鼻唇区域方面的度量精度。使用 3D FS 以及 IOS 和以前验证过的固定 3D 扫描仪作为参考,对 10 个 UCLP 模型的鼻唇区域进行扫描。口内扫描直接在 UCLP 模型上进行。为了将 FS 应用于模型,将它们嵌入到 3D 打印的样本面部中。通过应用基于部分的最佳拟合算法,将两个测试组与参考对齐。然后使用 3D 分析工具对参考的度量偏差进行后续分析。平均距离和积分距离作为表面和体积比较的主要参数。点比较作为附加参数。使用独立样本 t 检验进行统计分析。考虑到平均距离和积分距离是扫描仪准确性 3D 评估的主要参数,与参考相比,FS 和 IOS 在扫描裂隙模型的度量精度方面存在显著差异。与以前描述的固定 3D 扫描仪作为参考和验证基线相比,IOS 被证明在扫描精度方面明显更准确。通过添加以前验证过的 ATOS III Triple Scan 蓝光扫描仪作为参考,进一步验证了所测试的 IOS 和 FS 对鼻唇区域的 3D 评估。与验证基线扫描相比,IOS 在实验性裂隙模型扫描中显示出更高的度量精度。所收集的数据为 IOS 在鼻唇区域的 3D 评估中的临床应用提供了基础。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/655b/10374634/4816d929a6e6/41598_2023_39352_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验