用于监测镥-前列腺特异性膜抗原(Lu-PSMA)治疗的360°碲锌镉(CZT)相机性能研究

Performance study of a 360° CZT camera for monitoring Lu-PSMA treatment.

作者信息

Vergnaud Laure, Badel Jean-Noël, Giraudet Anne-Laure, Kryza David, Mognetti Thomas, Baudier Thomas, Rida Hanan, Dieudonné Arnaud, Sarrut David

机构信息

Centre de lutte contre le cancer Léon Bérard, Lyon, France.

CREATIS, CNRS UMR 5220, INSERM U 1044, Université de Lyon, INSA-Lyon, Université Lyon 1, Lyon, France.

出版信息

EJNMMI Phys. 2023 Sep 22;10(1):58. doi: 10.1186/s40658-023-00576-1.

Abstract

BACKGROUND

The aim of this study was to investigate the quantification performance of a 360° CZT camera for Lu-based treatment monitoring.

METHODS

Three phantoms with known Lu activity concentrations were acquired: (1) a uniform cylindrical phantom for calibration, (2) a NEMA IEC body phantom for analysis of different-sized spheres to optimise quantification parameters and (3) a phantom containing two large vials simulating organs at risk for tests. Four sets of reconstruction parameters were tested: (1) Scatter, (2) Scatter and Point Spread Function Recovery (PSFR), (3) PSFR only and (4) Penalised likelihood option and Scatter, varying the number of updates (iterations × subsets) with CT-based attenuation correction only. For each, activity concentration (ARC) and contrast recovery coefficients (CRC) were estimated as well as root mean square. Visualisation and quantification parameters were applied to reconstructed patient image data.

RESULTS

Optimised quantification parameters were determined to be: CT-based attenuation correction, scatter correction, 12 iterations, 8 subsets and no filter. ARC, CRC and RMS results were dependant on the methodology used for calculations. Two different reconstruction parameters were recommended for visualisation and for quantification. 3D whole-body SPECT images were acquired and reconstructed for Lu-PSMA patients in 2-3 times faster than the time taken for a conventional gamma camera.

CONCLUSION

Quantification of whole-body 3D images of patients treated with Lu-PSMA is feasible and an optimised set of parameters has been determined. This camera greatly reduces procedure time for whole-body SPECT.

摘要

背景

本研究的目的是调查一款360°碲锌镉(CZT)相机用于镥(Lu)基治疗监测的定量性能。

方法

采集了三个具有已知Lu活度浓度的体模:(1)一个用于校准的均匀圆柱形体模,(2)一个用于分析不同尺寸球体以优化定量参数的NEMA IEC体模,以及(3)一个包含两个模拟危险器官的大药瓶的体模用于测试。测试了四组重建参数:(1)散射校正,(2)散射校正和点扩散函数恢复(PSFR),(3)仅PSFR,以及(4)惩罚似然选项和散射校正,仅使用基于CT的衰减校正并改变更新次数(迭代次数×子集数)。对于每组参数,估计活度浓度(ARC)、对比度恢复系数(CRC)以及均方根。将可视化和定量参数应用于重建的患者图像数据。

结果

确定优化后的定量参数为:基于CT的衰减校正、散射校正、12次迭代、8个子集且不使用滤波器。ARC、CRC和RMS结果取决于用于计算的方法。推荐两种不同的重建参数用于可视化和定量分析。为Lu-PSMA患者采集并重建了3D全身SPECT图像,其速度比传统伽马相机快2至3倍。

结论

对接受Lu-PSMA治疗的患者进行全身3D图像定量分析是可行的,并且已经确定了一组优化参数。这款相机大大缩短了全身SPECT的检查时间。