Nuttens Victor, Schramm Georg, D'Asseler Yves, Koole Michel
Nuclear Medicine, OLV Aalst, Aalst, Belgium.
Nuclear Medicine and Molecular Imaging, KU Leuven, Leuven, Belgium.
EJNMMI Phys. 2024 Mar 19;11(1):29. doi: 10.1186/s40658-024-00627-1.
Next-generation SPECT/CT systems with CdZnTe (CZT) digital detectors in a ring-like setup are emerging to perform quantitative Lu-177 SPECT imaging in clinical routine. It is essential to assess how the shorter acquisition time might affect the image quality and uncertainty on the mean absorbed dose of the tumors and organs at risk compared to a conventional system.
A NEMA Image Quality phantom was scanned with a 3D CZT SPECT/CT system (Veriton, by Spectrum Dynamics) using 6 min per bed position and with a conventional SPECT/CT system (Symbia T16, by Siemens) using 16 min per bed position. The sphere-to-background ratio was 12:1 and the background activity concentration ranged from 0.52 to 0.06 MBq/mL. A clinical reconstruction protocol for dosimetry purposes was determined for both systems by maximizing the sphere-to-background ratio while keeping the coefficient of variation of the background as low as possible. The corresponding image resolution was determined by the matching filter method and used for a dose uncertainty assessment of both systems following an established uncertainty model..
The optimized iterative reconstruction protocol included scatter and attenuation correction for both systems and detector response modeling for the Siemens system. For the 3D CZT system, 6 iterations and 8 subsets were combined with a Gaussian post-filter of 3 mm Full Width Half Maximum (FWHM) for post-smoothing. For the conventional system, 16 iterations and 16 subsets were applied with a Gaussian post-smoothing filter of 1 mm FWHM. For these protocols, the sphere-to-background ratio was 18.5% closer to the true ratio for the conventional system compared to the 3D CZT system when considering the four largest spheres. Meanwhile, the background coefficient of variation was very similar for both systems. These protocols resulted in SPECT image resolution of 14.8 mm and 13.6 mm for the 3D CZT and conventional system respectively. Based on these resolution estimates, a 50% dose uncertainty corresponded to a lesion volume of 28 mL for the conventional system and a lesion volume of 33 mL for the 3D CZT system.
An optimized reconstruction protocol for a Veriton system with 6 min of acquisition time per bed position resulted in slightly higher dose uncertainties than a conventional Symbia system using 16 min of acquisition time per bed position. Therefore, a 3D CZT SPECT/CT allows to significantly reduce the acquisition times with only a very limited impact on dose uncertainties such that quantitative Lu-177 SPECT/CT imaging becomes much more accessible for treatment concurrent dosimetry. Nevertheless, the uncertainty of SPECT-based dose estimates remains high.
配备碲锌镉(CZT)数字探测器的环形设置的新一代SPECT/CT系统正在兴起,以在临床常规中进行定量Lu-177 SPECT成像。与传统系统相比,评估较短的采集时间如何影响图像质量以及肿瘤和危及器官的平均吸收剂量的不确定性至关重要。
使用3D CZT SPECT/CT系统(Veriton,Spectrum Dynamics公司)以每个床位6分钟的时间扫描NEMA图像质量体模,使用传统SPECT/CT系统(Symbia T16,西门子公司)以每个床位16分钟的时间进行扫描。球与背景的比值为12:1,背景活度浓度范围为0.52至0.06 MBq/mL。通过最大化球与背景的比值同时保持背景的变异系数尽可能低,为两个系统确定了用于剂量测定目的的临床重建方案。通过匹配滤波法确定相应的图像分辨率,并根据既定的不确定性模型用于两个系统的剂量不确定性评估。
优化的迭代重建方案包括两个系统的散射和衰减校正以及西门子系统的探测器响应建模。对于3D CZT系统,6次迭代和8个子集与3毫米半高宽(FWHM)的高斯后滤波器相结合用于后平滑处理。对于传统系统,应用16次迭代和16个子集以及1毫米FWHM的高斯后平滑滤波器。对于这些方案,在考虑四个最大球体时,与3D CZT系统相比,传统系统的球与背景的比值更接近真实比值18.5%。同时两个系统的背景变异系数非常相似。这些方案分别使3D CZT系统和传统系统的SPECT图像分辨率达到14.8毫米和13.6毫米。基于这些分辨率估计,50%的剂量不确定性对应于传统系统中28 mL的病变体积和3D CZT系统中33 mL的病变体积。
每个床位采集时间为6分钟的Veriton系统的优化重建方案导致的剂量不确定性略高于每个床位采集时间为16分钟的传统Symbia系统。因此,3D CZT SPECT/CT能够显著减少采集时间,对剂量不确定性的影响非常有限,使得定量Lu-177 SPECT/CT成像在治疗同期剂量测定中更容易实现。然而,基于SPECT的剂量估计的不确定性仍然很高。