Pulkki K, Irjala K
Scand J Clin Lab Invest. 1986 Oct;46(6):605-7. doi: 10.3109/00365518609083720.
Serum C-reactive protein (CRP) concentrations of 708 patients with suspected bacterial disease were estimated with a semiquantitative CRP latex slide test using cut-off levels of 20 and 40 mg/l and with quantitative turbidimetric technique. Latex slide tests gave 11.8% (80/708) discordant results. About 7% of tests gave clear disconcordant results (5% of those were positive and 2% negative tests). We conclude that CRP latex slide test is not useful in emergency laboratory in hospital material with a high incidence of bacterial infections.
采用截断值为20和40mg/l的半定量CRP乳胶玻片试验及定量比浊法技术,对708例疑似细菌性疾病患者的血清C反应蛋白(CRP)浓度进行了测定。乳胶玻片试验得出了11.8%(80/708)的不一致结果。约7%的检测给出了明显不一致的结果(其中5%为阳性检测,2%为阴性检测)。我们得出结论,CRP乳胶玻片试验在医院中细菌感染发生率较高的急诊实验室中并无用处。