Richards N P, Elliott T S, Powell R J, O'Callaghan C, Franklin P
J Clin Pathol. 1985 Apr;38(4):464-7. doi: 10.1136/jcp.38.4.464.
Serum C reactive protein concentrations measured by a laboratory based assay were compared with the semiquantitative results obtained with a visual agglutination method (Well-cotest, CRP kit). Using this agglutination kit, diluting serum 1/10 and 1/20 gave C reactive protein results which could be of more clinical value than those obtained using the 1/2, 1/4, 1/8, and 1/16 dilutions recommended by the manufacturers. The kit was also used on the ward by junior medical staff, who showed that after minimal training reproducible serum C reactive protein results could be obtained.
将通过实验室检测方法测得的血清C反应蛋白浓度与采用视觉凝集法(Well-cotest,CRP试剂盒)获得的半定量结果进行比较。使用该凝集试剂盒,将血清稀释1/10和1/20时得到的C反应蛋白结果可能比制造商推荐的1/2、1/4、1/8和1/16稀释度所获得的结果更具临床价值。该试剂盒也由初级医务人员在病房使用,他们表明经过最少的培训就能获得可重复的血清C反应蛋白结果。