Cascos Rocío, Celemín-Viñuela Alicia, Mory-Rubiños Nataly, Gómez-Polo Cristina, Ortega Rocío, Agustín-Panadero Rubén, Gómez-Polo Miguel
Department of Conservative Dentistry and Orofacial Prosthodontics, Faculty of Dentistry, Complutense University of Madrid, 28040 Madrid, Spain.
Department of Nursing and Estomatology, Faculty of Health Sciences, Rey Juan Carlos University, 28922 Madrid, Spain.
Materials (Basel). 2023 Oct 1;16(19):6532. doi: 10.3390/ma16196532.
This in vitro study aimed to assess the presence of microgaps at the implant-abutment interface in monolithic zirconia partial implant-supported fixed prostheses on transepithelial abutments versus Ti-base abutments.
Sixty conical connection dental implants were divided into two groups ( = 30). The control group consisted of three-unit bridge monolithic zirconia connected to two implants by a transepithelial abutment. The test group consisted of monolithic zirconia three-unit restoration connected to two implants directly by a titanium base (Ti-base) abutment. The sample was subjected to thermocycling (10,000 cycles at 5 °C to 55 °C, dwelling time 50 s) and chewing simulation (300,000 cycles, under 200 N at frequencies of 2 Hz, at a 30° angle). The microgap was evaluated at six points (mesiobuccal, buccal, distobuccal, mesiolingual, lingual, and distolingual) of each implant-abutment interface by using a scanning electron microscope (SEM). The data were analyzed using the Mann-Whitney tests ( > 0.05).
The SEM analysis showed a smaller microgap at the implant-abutment interface in the control group (0.270 μm) than in the test group (3.902 μm). Statistically significant differences were observed between both groups ( < 0.05).
The use or not of transepithelial abutments affects the microgap size. The transepithelial abutments group presented lower microgap values at the interface with the implant than the Ti-base group in monolithic zirconia partial implant-supported fixed prostheses. However, both groups had microgap values within the clinically acceptable range.
本体外研究旨在评估在经上皮基台与钛基台的单块氧化锆部分种植体支持的固定修复体中,种植体-基台界面处微间隙的存在情况。
60颗锥形连接牙种植体被分为两组(每组n = 30)。对照组由通过经上皮基台连接到两颗种植体的三单位桥单块氧化锆组成。试验组由通过钛基(Ti基)基台直接连接到两颗种植体的单块氧化锆三单位修复体组成。样本进行热循环(在5℃至55℃下循环10,000次,停留时间50秒)和咀嚼模拟(300,000次循环,在200N力、2Hz频率、30°角度下)。使用扫描电子显微镜(SEM)在每个种植体-基台界面的六个点(近中颊侧、颊侧、远中颊侧、近中舌侧、舌侧和远中舌侧)评估微间隙。使用Mann-Whitney检验分析数据(P > 0.05)。
SEM分析显示,对照组种植体-基台界面的微间隙(0.270μm)小于试验组(3.902μm)。两组之间观察到统计学上的显著差异(P < 0.05)。
是否使用经上皮基台会影响微间隙大小。在单块氧化锆部分种植体支持的固定修复体中,经上皮基台组在与种植体的界面处的微间隙值低于钛基组。然而,两组的微间隙值均在临床可接受范围内。