Agreement and Reproducibility of Anterior Chamber Angle Measurements between CASIA2 Built-In Software and Human Graders.

作者信息

Espinoza Gustavo, Iglesias Katheriene, Parra Juan C, Rodriguez-Una Ignacio, Serrano-Gomez Sergio, Prada Angelica M, Galvis Virgilio

机构信息

Centro Oftalmológico Virgilio Galvis, Floridablanca 681004, Santander, Colombia.

Fundación Oftalmológica de Santander, Floridablanca 681004, Santander, Colombia.

出版信息

J Clin Med. 2023 Oct 6;12(19):6381. doi: 10.3390/jcm12196381.

Abstract

PURPOSE

This study evaluated the agreement and reproducibility of ACA measurements obtained using the built-in software of the CASIA2 (Version 3G.1) and the measurements derived from expert clinicians.

METHODS

Healthy volunteers underwent ophthalmological evaluation and AS-OCT examination. ACA measurements derived from automated and manual SS location were obtained using the CASIA2 automated software and clinician identification, respectively. The intraobserver, interobserver reproducibility, CASIA2-human grader reproducibility and CASIA2 repeatability were assessed using intraclass correlation coefficients (ICCs).

RESULTS

The study examined 58 eyes of 30 participants. The CASIA2 software showed excellent repeatability for all ACA parameters (ICC > 0.84). Intraobserver, interobserver, and CASIA2-human grader reproducibility were also excellent (ICC > 0.87). Interobserver agreement was high, except for nasal TISA500, differing between observers 1 and 2 ( < 0.05). The agreement between CASIA2 measurements and human graders was high, except for nasal TISA500, where observer 1 values were smaller ( < 0.05).

CONCLUSION

The CASIA2 built-in software reliably measures ACA parameters in healthy individuals, demonstrating high consistency. Although a small difference was observed in nasal TISA500 measurements, interobserver and CASIA2-human grader reproducibility remained excellent. Automated SS detection has the potential to facilitate evaluation and monitoring of primary angle closure disease.

摘要

目的

本研究评估了使用CASIA2(版本3G.1)内置软件获得的前房角(ACA)测量值与专家临床医生测量值之间的一致性和可重复性。

方法

健康志愿者接受眼科评估和眼前节光学相干断层扫描(AS-OCT)检查。分别使用CASIA2自动化软件和临床医生识别,从自动和手动的巩膜嵴(SS)定位获得ACA测量值。使用组内相关系数(ICC)评估观察者内、观察者间的可重复性、CASIA2与人工分级者的可重复性以及CASIA2的重复性。

结果

该研究检查了30名参与者的58只眼睛。CASIA2软件对所有ACA参数均显示出出色的重复性(ICC>0.84)。观察者内、观察者间以及CASIA2与人工分级者的可重复性也都非常出色(ICC>0.87)。除鼻侧TISA500外,观察者间的一致性较高,观察者1和观察者2之间存在差异(<0.05)。除鼻侧TISA500外,CASIA2测量值与人工分级者之间的一致性较高,观察者1的测量值较小(<0.05)。

结论

CASIA2内置软件能够可靠地测量健康个体的ACA参数,显示出高度的一致性。尽管在鼻侧TISA500测量中观察到了微小差异,但观察者间以及CASIA2与人工分级者的可重复性仍然出色。自动SS检测有潜力促进原发性闭角型青光眼的评估和监测。