Anterior chamber angle imaging with swept-source optical coherence tomography: comparison between CASIAII and ANTERION.

机构信息

Department of Ophthalmology & Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, Kowloon, Hong Kong, People's Republic of China.

Hong Kong Eye Hospital, Hong Kong, SAR, People's Republic of China.

出版信息

Sci Rep. 2020 Oct 30;10(1):18771. doi: 10.1038/s41598-020-74813-3.

Abstract

This study compared the test-retest variabilities and measurement agreement of anterior chamber angle (ACA) dimensions measured by two anterior segment swept-source optical coherence tomography (SS-OCT)-the ANTERION (Heidelberg Engineering, Heidelberg, Germany) and CASIAII (Tomey, Nagoya, Japan). Thirty-eight subjects, 18 patients with primary angle closure and 20 healthy participants with open angles, were included. The mean age was 54.7 ± 15.8 years (range: 26-75 years). One eye of each subject was randomly selected for anterior segment imaging by ANTERION and CASIAII, using the same scan pattern (6 evenly spaced radial scans across the anterior segment for three times) in the same visit. The between- and within-instrument agreement and repeatability coefficients of angle open distance (AOD500), trabecular-iris space area (TISA500), lens vault (LV), scleral spur-scleral spur distance (SSD), anterior chamber depth (ACD), and pupil diameter (PD) were measured. The anterior and posterior boundaries of the cornea, iris, and lens were automatically segmented by the SS-OCT instruments; the scleral spur was manually located by a single masked observer. There were significant differences between ANTERION and CASIAII measurements; the SSD, PD, and ACD were smaller whereas AOD500 and TISA500 were greater in ANTERION compared with CASIAII (P < 0.001). Anterior segment measurements obtained from the two SS-OCT instruments showed strong associations (R ranged between 0.866 and 0.998) although the between-instrument agreement was poor; the spans of 95% limits of between-instrument agreement were ≥ 1.5-folds than the within-instrument agreement for either instrument. Whereas both SS-OCT instruments showed low test-retest measurement variabilities, the repeatability coefficients of AOD500, TISA500, ACD, and PD were slightly smaller for CASIAII than ANTERION (P ≤ 0.012).

摘要

本研究比较了两种前节扫频源光学相干断层扫描仪(SS-OCT)——ANTERION(德国海德堡工程公司)和 CASIAII(日本富美来)测量前房角(ACA)参数的重复性和测量一致性。共纳入 38 名受试者,18 名原发性闭角型青光眼患者和 20 名健康的开角型青光眼患者。平均年龄为 54.7±15.8 岁(26-75 岁)。每位受试者的每只眼均随机选择使用 ANTERION 和 CASIAII 进行前节成像,在同一次就诊时使用相同的扫描模式(前节 6 个均匀间隔的径向扫描,共 3 次)。测量角间隙距离(AOD500)、小梁虹膜空间面积(TISA500)、晶状体悬韧带(LV)、巩膜突巩膜突距离(SSD)、前房深度(ACD)和瞳孔直径(PD)的仪器间和仪器内一致性和重复性系数。SS-OCT 仪器自动对角膜、虹膜和晶状体的前、后边界进行分割;由一位单独的、经过训练的观察者手动定位巩膜突。与 CASIAII 相比,ANTERION 的 SSD、PD 和 ACD 较小,而 AOD500 和 TISA500 较大(P<0.001)。两种 SS-OCT 仪器获得的前节测量值具有很强的相关性(R 值介于 0.866 和 0.998 之间),尽管仪器间的一致性较差;两种仪器的仪器间一致性的 95%置信区间跨度均大于仪器内一致性的 1.5 倍。虽然两种 SS-OCT 仪器的测试-再测试测量变异性均较低,但 CASIAII 的 AOD500、TISA500、ACD 和 PD 的重复性系数略小于 ANTERION(P≤0.012)。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a824/7603499/d14297c33659/41598_2020_74813_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验