使用CASIA2和Spectralis OCT2光学相干断层扫描设备获得的眼前节参数测量的可重复性和一致性

Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography Devices.

作者信息

Xu Benjamin Y, Mai Derek D, Penteado Rafaella C, Saunders Luke, Weinreb Robert N

机构信息

Department of Ophthalmology, Hamilton Glaucoma Center, Shiley Eye Institute, University of California, San Diego, La Jolla, CA.

出版信息

J Glaucoma. 2017 Nov;26(11):974-979. doi: 10.1097/IJG.0000000000000788.

Abstract

PURPOSE

To assess the reproducibility and agreement of measurement values obtained from the Tomey CASIA2 and Heidelberg Spectralis OCT2 anterior segment optical coherence tomographic devices.

METHODS

Twenty eyes from 10 subjects ranging from age 28 to 45 years with no history of eye conditions or intraocular surgery were included. Two scans were obtained with each device in a standardized dark room environment after a period of dark adaptation. One anterior segment optical coherence tomography image along the horizontal (temporal nasal) meridian was analyzed per eye and per scan. Lens vault, pupil diameter, anterior chamber width, angle opening distance, trabecular iris space area, and scleral spur angle were measured using manufacturer-provided image analysis programs. Intraclass correlation coefficient (ICC) values, coefficients of variation, and Bland-Altman plots were computed to assess the intradevice correlation and interdevice agreement of measurement values.

RESULTS

There was excellent intradevice reproducibility of measurement values for both the CASIA (ICC range, 0.86 to 0.99) and Spectralis (ICC range, 0.79 to 1.00). There was also excellent interdevice correlation of measurement values (ICC range, 0.78 to 0.93) for all parameters except anterior chamber width (ICC 0.20). Linear regression models and Bland-Altman plots showed that this relationship was strongest when measurement values were small.

CONCLUSIONS

There is excellent intradevice reproducibility and good interdevice agreement of anterior segment parameter measurement values for the CASIA2 and Spectralis OCT2. However, the measurements obtained with each device should not be considered interchangeable.

摘要

目的

评估从Tomey CASIA2和海德堡Spectralis OCT2眼前节光学相干断层扫描设备获得的测量值的可重复性和一致性。

方法

纳入10名年龄在28至45岁之间、无眼部疾病或眼内手术史的受试者的20只眼睛。在标准化暗室环境中经过一段时间的暗适应后,使用每个设备进行两次扫描。每只眼睛每次扫描分析一张沿水平(颞侧-鼻侧)子午线的眼前节光学相干断层扫描图像。使用制造商提供的图像分析程序测量晶状体拱高、瞳孔直径、前房宽度、房角开放距离、小梁虹膜间隙面积和巩膜突角度。计算组内相关系数(ICC)值、变异系数和Bland-Altman图,以评估测量值在设备内的相关性和设备间的一致性。

结果

CASIA(ICC范围为0.86至0.99)和Spectralis(ICC范围为0.79至1.00)的测量值在设备内均具有出色的可重复性。除前房宽度(ICC为0.20)外,所有参数的测量值在设备间也具有出色的相关性(ICC范围为0.78至0.93)。线性回归模型和Bland-Altman图显示,当测量值较小时,这种关系最强。

结论

CASIA2和Spectralis OCT2的眼前节参数测量值在设备内具有出色的可重复性,在设备间具有良好的一致性。然而,不应认为每个设备获得的测量值是可互换的。

相似文献

4
Repeatability and Agreement of Two Swept-Source Optical Coherence Tomographers for Anterior Segment Parameter Measurements.
J Glaucoma. 2022 Jul 1;31(7):602-608. doi: 10.1097/IJG.0000000000001989. Epub 2022 Jan 20.
6
Swept-source optical coherence tomography and ultrasound biomicroscopy study of anterior segment parameters in primary angle-closure glaucoma.
Graefes Arch Clin Exp Ophthalmol. 2023 Jun;261(6):1651-1658. doi: 10.1007/s00417-022-05970-6. Epub 2023 Jan 11.
7
Reproducibility and agreement of four anterior segment-optical coherence tomography devices for anterior chamber angle measurements.
Graefes Arch Clin Exp Ophthalmol. 2020 Jul;258(7):1475-1481. doi: 10.1007/s00417-020-04648-1. Epub 2020 Mar 26.
9
Anterior segment biometry using spectral-domain optical coherence tomography.
J Refract Surg. 2014 May;30(5):354-60. doi: 10.3928/1081597X-20140326-01. Epub 2014 Apr 2.
10
Assessment of anterior segment measurements using a high-resolution imaging device.
Expert Rev Med Devices. 2020 Sep;17(9):969-979. doi: 10.1080/17434440.2020.1816463. Epub 2020 Sep 7.

引用本文的文献

2
Pilot Study on Postoperative Toric Intraocular Lens Alignment: Comparing Casia2 and Photoshop Imaging Techniques.
J Ophthalmol. 2024 Nov 14;2024:1053914. doi: 10.1155/2024/1053914. eCollection 2024.
3
Performance of anterior segment OCT-based algorithms in the opportunistic screening for primary angle-closure disease.
Heliyon. 2024 Mar 30;10(7):e28885. doi: 10.1016/j.heliyon.2024.e28885. eCollection 2024 Apr 15.
5
Comparison of corneal and lens density measurements obtained by Pentacam and CASIA2 in myopes.
BMC Ophthalmol. 2023 Nov 10;23(1):448. doi: 10.1186/s12886-023-03199-3.
8
Comparison of a New Scheimpflug Camera and Swept-Source Optical Coherence Tomographer for Measurements of Anterior Segment Parameters.
Ophthalmol Ther. 2023 Dec;12(6):3187-3198. doi: 10.1007/s40123-023-00813-x. Epub 2023 Sep 25.
9
Improvement in the ideal range of vault after implantable collamer lens implantation: a new vault prediction formula.
Front Med (Lausanne). 2023 Apr 27;10:1132102. doi: 10.3389/fmed.2023.1132102. eCollection 2023.
10

本文引用的文献

1
Association of Baseline Anterior Segment Parameters With the Development of Incident Gonioscopic Angle Closure.
JAMA Ophthalmol. 2017 Mar 1;135(3):252-258. doi: 10.1001/jamaophthalmol.2016.5847.
5
Optimal number of angle images for calculating anterior angle volume and iris volume measurements.
Invest Ophthalmol Vis Sci. 2015 May;56(5):2842-7. doi: 10.1167/iovs.14-15883.
8
Longitudinal changes of angle configuration in primary angle-closure suspects: the Zhongshan Angle-Closure Prevention Trial.
Ophthalmology. 2014 Sep;121(9):1699-1705. doi: 10.1016/j.ophtha.2014.03.039. Epub 2014 May 15.
9
Anterior segment biometry using spectral-domain optical coherence tomography.
J Refract Surg. 2014 May;30(5):354-60. doi: 10.3928/1081597X-20140326-01. Epub 2014 Apr 2.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验