两种分光光度计测量 VITA 经典比色板和 VITA 3D Master 比色板的 Cie L*A*B*值。

Cie L*A*B* Values in VITA Classical and VITA 3D Master by Two Dental Spectrophotometers.

出版信息

Int J Prosthodont. 2023 Nov 1;36(5):649. doi: 10.11607/ijp.7235.

Abstract

PURPOSE

To compare the CIE Lab* values of two different dental shade guides using two shade-matching spectrophotometers.

MATERIALS AND METHODS

SpectroShade Micro (SS) and Easyshade (ES) (VITA Zahnfabrik) were tested. One calibrated operator performed 30 measurements of each tab from three batches of the two guides, VITA Classical (VC) and VITA 3D Shade Master (VM). Receiver operating characteristic (ROC) curves and intraclass coefficients (ICC) between the different batches were calculated. Results were given as mean and SD of the Lab* values and the respective color differences according to the CIEDE 2000 formula (ΔE00) for each of the shade tabs in the two spectrophotometers and then analyzed with independent student t test (α = .05).

RESULTS

A total of 1,440 and 2,610 measurements for VC and VM, respectively, were performed per spectrophotometer. There were statistically significant differences between the two devices for all Lab* values, with the exception of: L* for 3R1.5; a* for 2R1.5, 3L1.5, 2L2.5, 3M2, 3L2.5, 4L1.5, and 4L2.5; and b* for D3, 1M2, and 3M2. When assessing the same shade guide, differences in ΔE00 were detected between devices above the acceptability threshold (AT) (ΔE00 ≥ 1.8) for all shade tabs except for VC (C1, C2, D3, A3.5, C3, and A4) and VM (4M1, 3R2.5, 4L2.5, and 4M3). The overall mean of the interdevice ΔE00 was 2.2 ± 1.0 for VC and 2.5 ± 1.0 for VM.

CONCLUSIONS

The two dental spectrophotometers presented high ICC and ROC values, which validates their indication as auxiliary tools. However, there are discrepancies in the hues D (VC) and R (VM) with interdevice variability when evaluated for the Lab* component. Int J Prosthodont 2023;36:e38-e52.

摘要

目的

使用两种比色分光光度计比较两种不同牙科比色指南的 CIE Lab*值。

材料和方法

测试了 SpectroShade Micro(SS)和 Easyshade(ES)(VITA Zahnfabrik)。一名经过校准的操作员对来自两个指南(VITA Classical(VC)和 VITA 3D Shade Master(VM))的三个批次的每个标签进行了 30 次测量。计算了不同批次之间的接收者操作特征(ROC)曲线和组内系数(ICC)。结果以每个分光光度计中两个比色指南的每个色标 tab 的 Lab*值和相应的 CIEDE 2000 公式(ΔE00)的色差的平均值和标准差给出,然后使用独立学生 t 检验进行分析(α=0.05)。

结果

每个分光光度计分别对 VC 和 VM 进行了总共 1440 和 2610 次测量。两种设备之间的所有 Lab值均存在统计学差异,除了:L值为 3R1.5;a值为 2R1.5、3L1.5、2L2.5、3M2、3L2.5、4L1.5 和 4L2.5;b值为 D3、1M2 和 3M2。当评估同一比色指南时,除 VC(C1、C2、D3、A3.5、C3 和 A4)和 VM(4M1、3R2.5、4L2.5 和 4M3)外,设备之间的ΔE00 差异检测到超出可接受阈值(AT)(ΔE00≥1.8)。两种设备的总体平均 interdeviceΔE00 分别为 VC 为 2.2±1.0,VM 为 2.5±1.0。

结论

两种牙科分光光度计的 ICC 和 ROC 值均较高,这验证了它们作为辅助工具的使用。然而,在评估 Lab*成分时,设备之间存在 D(VC)和 R(VM)色调的差异,具有变异性。