Center for Reproductive Medicine, Zhongda Hospital, Southeast University, Nanjing, Jiangsu, China.
J Clin Lab Anal. 2023 Dec;37(23-24):e24986. doi: 10.1002/jcla.24986. Epub 2023 Nov 27.
The performance evaluation of each computer-assisted sperm analysis (CASA) system may provide a basis for the interpretation of clinical results and further improvement of the CASA system.
The accuracy of the GSA-810 CASA system was evaluated by detecting latex bead quality control products. The precision of sperm concentration, morphology, and percentages of progressively motile sperm (PR) were evaluated by coefficient of variation (CV). Three samples with sperm concentration of about 100 × 10 /mL were diluted to evaluate the linear range.
The detection values of latex beads were within the range of target values. The CVs of sperm concentration and PR were significantly and negatively correlated with sperm concentration (r = -0.561, p = 0.001) and PR value (r = -0.621, p < 0.001), respectively. The R values of the linear range of sperm concentration were ≥0.99. There was no significant difference in sperm motility and PR within 1-10 min at 36.5°C ± 0.5°C. The coincidence rates of sperm morphology and sperm head morphology for 36 semen samples analyzed by the GSA-810 system and manual method were 99.40% and 99.67%, respectively. The CVs of the percentage of sperm with abnormal morphology and percentage of sperm with abnormal head morphology were less than 5%.
The GSA-810 system can accurately analyze normal semen samples, but the repeatability of the results is poor for oligozoospermia and asthenozoospermia samples. The future CASA system for analyzing sperm morphology should focus on recognizing the middle and tail segments of a spermatozoon.
每个计算机辅助精液分析(CASA)系统的性能评估可为临床结果的解释和进一步改进 CASA 系统提供依据。
通过检测乳胶珠质量控制品评估 GSA-810 CASA 系统的准确性。通过变异系数(CV)评估精子浓度、形态和前向运动精子(PR)百分比的精密度。将浓度约为 100×10/ml 的三个样本稀释,以评估线性范围。
乳胶珠的检测值在目标值范围内。精子浓度和 PR 的 CV 与精子浓度(r=-0.561,p=0.001)和 PR 值(r=-0.621,p<0.001)呈显著负相关。精子浓度线性范围的 R 值均≥0.99。在 36.5°C±0.5°C 下,1-10 分钟内精子活力和 PR 无显著差异。用 GSA-810 系统和手工法分析 36 份精液样本的精子形态和精子头形态的符合率分别为 99.40%和 99.67%。精子形态异常和精子头形态异常的百分比的 CV 值均小于 5%。
GSA-810 系统能够准确分析正常精液样本,但对少精症和弱精症样本的结果重复性较差。未来用于分析精子形态的 CASA 系统应重点识别精子的中段和尾部。