一种新型一体式光学生物测量仪的评估及其与经验证的扫频源 OCT 生物测量仪的比较。

Evaluation of a New All-in-One Optical Biometer and Comparison With a Validated Swept-source OCT Biometer.

出版信息

J Refract Surg. 2023 Dec;39(12):825-830. doi: 10.3928/1081597X-20231018-02. Epub 2023 Dec 1.

Abstract

PURPOSE

To assess agreement between a new all-in-one non-contact optical biometer based on optical low coherence reflectometry (SW-9000 µm Plus; Suoer) and a swept-source optical coherence tomography biometer (OA-2000; Tomey).

METHODS

Each eye was scanned three times in a row by each device at random. The measured ocular parameters included central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), axial length (AL), flat keratometry (Kf), steep keratometry (Ks), mean keratometry (Km), astigmatism, corneal diameter (CD), and pupil diameter (PD). The paired test was used to show the differences between the SW-9000 and OA-2000. Bland-Altman plots and the 95% limits of agreement (LoA) were applied to assess the consistency of the measurements.

RESULTS

Sixty eyes from 60 healthy participants were examined, with a mean spherical equivalent refraction of -5.58 ± 2.31 diopters and a mean age of 30.40 ± 6.07 years. The Bland-Altman plots showed high agreement for AL, ACD, LT, Kf, Ks, Km, astigmatism, and CD measurements (95% LoA: -0.06 to 0.04 mm, -0.10 to 0.06 mm, -0.12 to 0.11 mm, -0.30 to 0.29 D, -0.35 to 0.38 D, -0.29 to 0.30 D, -0.30 to 0.34 D, and -0.50 to 0.06 mm, respectively), whereas the agreement for CCT and PD were moderate (95% LoA: 7.12 to 20.43 µm, -0.75 to 1.19 mm, respectively).

CONCLUSIONS

The new all-in-one non-contact biometer had high agreement with the OA-2000 biometer on the AL, ACD, LT, Kf, Ks, Km, astigmatism, and CD measurements. For most of the ocular parameters assessed, they were clinically interchangeable. .

摘要

目的

评估基于光学低相干反射计的新型一体式非接触式光学生物测量仪(SW-9000 µm Plus;索埃)与扫频源光相干断层生物测量仪(OA-2000;Tomey)之间的一致性。

方法

每个设备连续三次随机对每只眼睛进行扫描。所测量的眼部参数包括中央角膜厚度(CCT)、前房深度(ACD)、晶状体厚度(LT)、眼轴长度(AL)、平角膜曲率(Kf)、陡角膜曲率(Ks)、平均角膜曲率(Km)、散光、角膜直径(CD)和瞳孔直径(PD)。采用配对 t 检验显示 SW-9000 和 OA-2000 之间的差异。应用 Bland-Altman 图和 95%一致性界限(LoA)评估测量值的一致性。

结果

对 60 名健康参与者的 60 只眼进行了检查,平均等效球镜屈光度为-5.58 ± 2.31 屈光度,平均年龄为 30.40 ± 6.07 岁。Bland-Altman 图显示,AL、ACD、LT、Kf、Ks、Km、散光和 CD 测量值的一致性较高(95%LoA:-0.06 至 0.04mm、-0.10 至 0.06mm、-0.12 至 0.11mm、-0.30 至 0.29D、-0.35 至 0.38D、-0.29 至 0.30D、-0.30 至 0.34D 和-0.50 至 0.06mm),而 CCT 和 PD 的一致性为中度(95%LoA:7.12 至 20.43µm、-0.75 至 1.19mm)。

结论

新型一体式非接触式生物测量仪与 OA-2000 生物测量仪在 AL、ACD、LT、Kf、Ks、Km、散光和 CD 测量值上具有高度一致性。对于评估的大多数眼部参数,它们在临床上是可互换的。