比较新型扫频源光学相干断层扫描仪和普通光学低相干反射仪的眼部生物测量值。

Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry.

机构信息

School of Ophthalmology and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China.

Key Laboratory of Vision Science, Ministry of Health P.R. China, Wenzhou, Zhejiang, China.

出版信息

Sci Rep. 2017 May 30;7(1):2484. doi: 10.1038/s41598-017-02463-z.

Abstract

The purpose of the current study was to compare the measurements between a new optical biometer based on swept-source optical coherence tomography (SS-OCT), the OA-2000 (Tomey, Japan), and an optical biometer based on optical low coherence reflectometry (OLCR), the Lenstar (Haag-Streit, Switzerland). Ninety-nine eyes of 99 healthy subjects were included. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), keratometry (K) readings, including flat K (Kf), steep K (Ks), mean K (Km), astigmatism vectors J, J at diameters of 2.5 and 3.0 mm, and white-to-white diameter (WTW) were measured three times each using both biometer in normal eyes by random sequence. Bland-Altman analysis showed good agreement between the SS-OCT and OLCR devices for AL, AD, ACD, LT, with narrow 95% LoA (-0.05 to 0.07 mm, -0.09 to 0.10 mm, -0.10 to 0.09 mm, and -0.06 to 0.22 mm, respectively), and the P values of ACD were both >0.05. The CCT, Kf, Ks, Km, J, J and WTW values provided by the OA-2000 were in good agreement with the Lenstar, and statistically significant differences were detected for some of them but not clinical differences. The agreement was excellent especially for AL.

摘要

本研究的目的是比较基于扫频源光相干断层扫描(SS-OCT)的新型光学生物测量仪(OA-2000,Tomey,日本)和基于光低相干反射计(OLCR)的光学生物测量仪(Lenstar,Haag-Streit,瑞士)的测量结果。纳入了 99 名健康受试者的 99 只眼。分别用两种生物测量仪随机顺序测量了正常眼的眼轴长度(AL)、中央角膜厚度(CCT)、前房深度(ACD)、房水深度(AD)、晶状体厚度(LT)、角膜曲率读数(K),包括平 K(Kf)、陡 K(Ks)、平均 K(Km)、散光矢量 J、直径为 2.5 和 3.0mm 的 J 以及白到白直径(WTW),每种参数均测量 3 次。 Bland-Altman 分析显示,SS-OCT 和 OLCR 设备在 AL、AD、ACD、LT 方面具有良好的一致性,95%的 LOA 较窄(分别为-0.05 至 0.07mm、-0.09 至 0.10mm、-0.10 至 0.09mm 和-0.06 至 0.22mm),且 ACD 的 P 值均>0.05。OA-2000 提供的 CCT、Kf、Ks、Km、J、J 和 WTW 值与 Lenstar 一致,且其中一些值存在统计学差异,但无临床差异。尤其是 AL 方面,一致性非常好。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/eff7/5449389/4738656f03bb/41598_2017_2463_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验