Continuous fiberoptic arterial oxygen tension measurements in dogs.

作者信息

Barker S J, Tremper K K, Hyatt J, Zaccari J, Heitzmann H A, Holman B M, Pike K, Ring L S, Teope M, Thaure T B

出版信息

J Clin Monit. 1987 Jan;3(1):48-52. doi: 10.1007/BF00770884.

Abstract

An experimental study using a new fiberoptic sensor for the continuous intraarterial measurement of oxygen tension is described. This "optode" sensor uses the phenomenon of fluorescence quenching to determine the oxygen tension of the surrounding medium. To assess the accuracy of this device, we anesthetized 4 dogs and monitored them continuously with arterial catheters and an intraarterial optode probe, and intermittently with arterial blood gas analysis. The inspired oxygen fraction was varied from 1.0 to 0.1, and arterial blood gases were measured for comparison with the optode reading. Two hundred ninety data sets yielded a correlation coefficient of 0.96, with a linear regression slope of 0.98 and intercept of 5.1 mm Hg. In the 72 data sets from the last dog, the bias and precision of the optode arterial oxygen tension values were -10.3 mm Hg and 20.0 mm Hg, respectively. The optode probe was easily inserted through a 20-gauge catheter and did not interfere with continuous arterial pressure measurement or blood sampling. This study suggests that the optode has great potential as a continuous, real-time monitor of arterial oxygen tension.

摘要

本文描述了一项使用新型光纤传感器进行动脉内氧分压连续测量的实验研究。这种“光极”传感器利用荧光猝灭现象来测定周围介质的氧分压。为评估该装置的准确性,我们对4只狗进行麻醉,通过动脉导管和动脉内光极探头对其进行连续监测,并通过动脉血气分析进行间歇性监测。吸入氧分数从1.0变化至0.1,并测量动脉血气以与光极读数进行比较。290个数据集的相关系数为0.96,线性回归斜率为0.98,截距为5.1毫米汞柱。在最后一只狗的72个数据集中,光极动脉氧分压值的偏差和精密度分别为-10.3毫米汞柱和20.0毫米汞柱。光极探头可轻松通过20号导管插入,且不干扰动脉压的连续测量或血液采样。本研究表明,光极作为动脉氧分压的连续实时监测仪具有很大潜力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验