Kozłowski Filip M, van Reenen Christoffel J, Trauernicht Christoph J
Division of Medical Physics, Tygerberg Hospital, Cape Town, South Africa.
Department of Medical Physics, Faculty of Medicine and Health Sciences, Stellenbosch University, Cape Town, South Africa.
SA J Radiol. 2024 Jan 31;28(1):2809. doi: 10.4102/sajr.v28i1.2809. eCollection 2024.
Diagnostic reference levels (DRLs) are an important metric in identifying abnormally high radiation doses in diagnostic examinations. National DRLs for CT colonography do not currently exist in South Africa, but there are efforts to collect data for a national DRL project.
This study investigated radiation doses for CT colonography in adult patients at a large tertiary hospital in South Africa with the aim of setting local DRLs.
Patient data from two CT scanners (Philips Ingenuity and Siemens Somatom go.Top) in the period March 2020 - March 2023 were obtained from the hospital's picture archiving and communication system (PACS) ( = 115). Analysis involved determining the median computed tomography dose index-volume (CTDI) and dose-length product (DLP) values. The findings were compared with DRLs established internationally.
Ingenuity median CTDI was 20 mGy and DLP was 2169 mGycm; Somatom median CTDI was 6 mGy and DLP was 557 mGycm. Ingenuity exceeded the United Kingdom's (UK) recommended DRLs by 82% and 214%, respectively. Somatom median CTDI and DLP were 45% and 19% lower than UK NDRLs.
Somatom's tin filter and other dose reduction features provided significant dose reduction. These data were used to set DRLs for CT colonography at the hospital; CTDI: 6 mGy and DLP: 557 mGy*cm.
In addition to informing radiation protection practices at the level of the institution, the established local DRLs contribute towards implementing regional and national DRLs.
诊断参考水平(DRLs)是识别诊断检查中异常高辐射剂量的一项重要指标。南非目前尚无CT结肠成像的国家诊断参考水平,但正在努力为一项国家诊断参考水平项目收集数据。
本研究调查了南非一家大型三级医院成年患者CT结肠成像的辐射剂量,旨在设定当地的诊断参考水平。
从医院的图像存档与通信系统(PACS)获取2020年3月至2023年3月期间两台CT扫描仪(飞利浦Ingenuity和西门子Somatom go.Top)的患者数据(n = 115)。分析包括确定计算机断层扫描剂量指数-容积(CTDI)和剂量长度乘积(DLP)的中位数。将结果与国际上确定的诊断参考水平进行比较。
Ingenuity的CTDI中位数为20 mGy,DLP为2169 mGycm;Somatom的CTDI中位数为6 mGy,DLP为557 mGycm。Ingenuity分别比英国推荐的诊断参考水平高出82%和214%。Somatom的CTDI中位数和DLP分别比英国国家诊断参考水平低45%和19%。
Somatom的锡滤过器和其他剂量降低功能显著降低了剂量。这些数据被用于设定该医院CT结肠成像的诊断参考水平;CTDI:6 mGy,DLP:557 mGy*cm。
既定的当地诊断参考水平除了为机构层面的辐射防护实践提供信息外,还有助于实施区域和国家诊断参考水平。