Kim Byeong-Soo, Park Sam-Ho, Jung Sung-Soo, Kim Hong-Jun, Woo Seong-Dae, Lee Myung-Mo
Department of Physical Therapy, Daejeon University, Daejeon 34520, Republic of Korea.
Division of Pulmonary and Critical Care Medicine, Department of Internal Medicine, Chungnam National University Hospital, Daejeon 35015, Republic of Korea.
Healthcare (Basel). 2024 Feb 20;12(5):507. doi: 10.3390/healthcare12050507.
A spirometer is a medical device frequently used clinically for the diagnosis and prediction of lung disease. This study aimed to investigate the clinical usefulness of a hand-held spirometer (The Spirokit), compared with conventional spirometry in patients with chronic obstructive pulmonary disease (COPD). This study was conducted from February 2022 to October 2022. Measurements from 80 patients with COPD (male: 53, female: 27) were obtained using The Spirokit and PC-based pulmonary function test equipment, and the resulting values were compared and analyzed. For the concurrent validity comparison of The Spirokit, the intra-class correlation (ICC 2, 1), coefficients of variation (CV), 95% limits of agreement (95% LOA), and Cohen's Kappa Index were analyzed. The Spirokit showed high agreement (ICC: 0.929-0.989; 95% LOA: -0.525 to 2.559; and CV: 0.05-0.08) with the PC-based pulmonary function tester. Using the Cohen's kappa coefficients, the device showed high sensitivity, specificity, and accuracy scores of Pa: 0.90, Pc: 0.52, and K: 0.79, respectively, indicating considerable agreement. The Spirokit, a portable pulmonary function test device, is a piece of equipment with high validity and portability, with high potential for replacing PC-based pulmonary function test equipment.
肺活量计是临床上常用于诊断和预测肺部疾病的医疗设备。本研究旨在调查手持式肺活量计(The Spirokit)与传统肺活量测定法相比在慢性阻塞性肺疾病(COPD)患者中的临床实用性。本研究于2022年2月至2022年10月进行。使用The Spirokit和基于个人电脑的肺功能测试设备获取了80例COPD患者(男性:53例,女性:27例)的测量数据,并对所得值进行了比较和分析。为了对The Spirokit进行同时效度比较,分析了组内相关系数(ICC 2,1)、变异系数(CV)、95%一致性界限(95% LOA)和科恩kappa指数。The Spirokit与基于个人电脑的肺功能测试仪显示出高度一致性(ICC:0.929 - 0.989;95% LOA:-0.525至2.559;CV:0.05 - 0.08)。使用科恩kappa系数,该设备分别显示出高灵敏度、特异性和准确性得分,Pa为0.90,Pc为0.52,K为0.79,表明一致性相当高。The Spirokit,一种便携式肺功能测试设备,是一种具有高有效性和便携性的设备,具有很高的潜力可替代基于个人电脑的肺功能测试设备。