Uncertainty Analysis of Time-Integrated Activity Coefficient in Single-Time-Point Dosimetry Using Bayesian Fitting Method.

作者信息

Jundi Achmad Faturrahman, Naqiyyun M Dlorifun, Patrianesha Bisma Barron, Mu'minah Intan A S, Riana Ade, Hardiansyah Deni

机构信息

Medical Physics and Biophysics, Physics Department, Faculty of Mathematics and Natural Sciences, Universitas Indonesia, Depok, 16424 Indonesia.

Research Center for Safety, Metrology, and Nuclear Quality Technology, National Research and Innovation Agency, KST B. J. Habibie, South Tangerang, 15314 Indonesia.

出版信息

Nucl Med Mol Imaging. 2024 May;58(3):120-128. doi: 10.1007/s13139-024-00851-8. Epub 2024 Feb 23.

Abstract

PURPOSE

Calculation of the uncertainty of the individual time-integrated activity coefficient (TIACs) is desirable in molecular radiotherapy. However, the calculation of TIAC's uncertainty in single-time-point (STP) method has never been reported in the literature. This study presents a method based on the Bayesian fitting (BF) to calculate the standard deviation (SD) of individual TIACs in the STP dosimetry.

METHODS

Biokinetic data of Lu-DOTATATE in kidneys were obtained from PMID33443063. BF methods with extended objective function, which optimize the fitting using prior knowledge of the function's parameters, were used. Reference TIACs (rTIACs) were calculated by fitting a mono-exponential function to the all-time-point data. The goodness of fit was checked based on the visual inspection and the coefficient of variations (CV) of the fitted parameters < 0.5. BF with relative (BFr) and absolute-based (BFa) variance methods were used to obtain the calculated TIACs (cTIACs) from the STP dosimetry. Performance of the STP method was obtained by calculating the relative deviation (RD) between cTIACs and rTIACs.

RESULTS

Visual inspection showed a good fit for all patients with CV of fitted parameters less than 50%. The mean ± SD of cTIAC's %RD were 7.0 ± 25.2 for BFr and 2.6 ± 8.9 for BFa. The range of %CV of the individual cTIAC's SD for BFr and BFa methods was 36-78% and 22-33%, respectively, while the %CV of the rTIAC SD was 0.8-49%.

CONCLUSION

We introduce the BF method to calculate the SD of individual TIACs in STP dosimetry. The presented method might be used as an alternative method for uncertainty analysis in STP dosimetry.

SUPPLEMENTARY INFORMATION

The online version contains supplementary material available at 10.1007/s13139-024-00851-8.

摘要

目的

在分子放射治疗中,计算个体时间积分活度系数(TIACs)的不确定度是很有必要的。然而,单时间点(STP)方法中TIAC不确定度的计算在文献中从未有过报道。本研究提出一种基于贝叶斯拟合(BF)的方法来计算STP剂量测定中个体TIACs的标准差(SD)。

方法

从PMID33443063获取肾脏中镥-奥曲肽的生物动力学数据。使用具有扩展目标函数的BF方法,该方法利用函数参数的先验知识优化拟合。通过将单指数函数拟合到所有时间点数据来计算参考TIACs(rTIACs)。基于目视检查和拟合参数的变异系数(CV)<0.5来检查拟合优度。使用相对(BFr)和基于绝对的(BFa)方差方法的BF从STP剂量测定中获得计算得到的TIACs(cTIACs)。通过计算cTIACs与rTIACs之间的相对偏差(RD)来获得STP方法的性能。

结果

目视检查显示所有患者的拟合良好,拟合参数的CV小于50%。BFr方法中cTIAC的%RD的平均值±SD为7.0±25.2,BFa方法为2.6±8.9。BFr和BFa方法中个体cTIAC的SD的%CV范围分别为36 - 78%和22 - 33%,而rTIAC SD的%CV为0.8 - 49%。

结论

我们引入BF方法来计算STP剂量测定中个体TIACs的SD。所提出的方法可作为STP剂量测定中不确定度分析的替代方法。

补充信息

在线版本包含可在10.1007/s13139 - 024 - 00851 - 8获取的补充材料。