ATGC Inc., King of Prussia, PA, United States.
Instanosis Inc., King of Prussia, PA, United States.
Clin Chem. 2024 Jul 5;70(7):978-986. doi: 10.1093/clinchem/hvae071.
Abuse of fentanyl and its analogs is a major contributor to the opioid overdose epidemic in the United States, but detecting and quantifying trace amounts of such drugs remains a challenge without resorting to sophisticated mass spectrometry-based methods.
A sensitive immunoassay with a sub-picogram limit of detection for fentanyl and a wide range of fentanyl analogs has been developed, using a novel high-affinity antibody fused with NanoLuc, a small-size luciferase that can emit strong and stable luminescence. When used with human urine samples, the assay has a sub-picogram limit of detection for fentanyl, with results fully concordant with LC-MS.
When applied to clinical samples, the novel chemiluminescence immunoassay can detect low positive fentanyl missed by routine screening immunoassays, with a limit of detection of 0.8 pg/mL in human urine. When applied to environmental samples, the assay can detect levels as low as 0.25 pg fentanyl per inch2 of environment surface. Assay turnaround time is less than 1 h, with inexpensive equipment and the potential for high-throughput automation or in-field screening.
We have established a novel assay that may have broad applications in clinical, environmental, occupational, and forensic scenarios for detection of trace amounts of fentanyl and its analogs.
芬太尼及其类似物的滥用是导致美国阿片类药物过量流行的主要原因,但如果不采用复杂的基于质谱的方法,检测和定量痕量此类药物仍然是一个挑战。
我们开发了一种针对芬太尼和广泛的芬太尼类似物的具有纳克级检测限的灵敏免疫测定法,该方法使用与 NanoLuc 融合的新型高亲和力抗体,NanoLuc 是一种小型荧光素酶,可发出强而稳定的荧光。当用于人类尿液样本时,该测定法对芬太尼的检测限为纳克级,结果与 LC-MS 完全一致。
当应用于临床样本时,新型化学发光免疫测定法可以检测到常规筛选免疫测定法漏检的低浓度阳性芬太尼,在人类尿液中的检测限为 0.8 pg/mL。当应用于环境样本时,该测定法可以检测到低至每平方英寸环境表面 0.25 皮克的芬太尼。测定法的周转时间不到 1 小时,使用的设备廉价,并且具有高通量自动化或现场筛查的潜力。
我们已经建立了一种新的测定法,它可能在临床、环境、职业和法医场景中广泛应用于检测痕量芬太尼及其类似物。