Department of Ophthalmology, Gachon University Gil Hospital, Incheon, Korea.
Department of Ophthalmology, Seoul National University College of Medicine, Seoul, South Korea.
Sci Rep. 2024 Oct 17;14(1):24368. doi: 10.1038/s41598-024-76031-7.
To evaluate the agreement of ocular biometry measured using a swept-source optical coherence (Casia 2) and a dual Scheimpflug (Galilei G6) tomography in keratoconus.
This retrospective study included 102 eyes from 102 keratoconus patient examined using both devices. Parameters compared included flat (Kf) and steep (Ks) keratometry, astigmatism of anterior, posterior, and total keratometry, central (CCT) and thinnest (TCT) corneal thickness. To assess the agreement, intraclass coefficient (ICC) and Bland-Altman analysis with 95% limits of agreement (LoA) were used.
Anterior and total Ks and Kf showed moderate or good agreement with 95% limits of agreement (LoA) range over 9.75 D. Posterior Ks and Kf were lower in the Galilei G6 (all p < 0.001). Astigmatism showed moderate, poor, and moderate agreement for anterior, posterior, and total keratometry, respectively. CCT and TCT showed excellent agreement; however, the 95% LoA range was over 60 μm.
The agreement between the two devices was not excellent for most parameters used to diagnose keratoconus and assess disease progression, and the differences were clinically significant. Therefore, the measurements from these two devices are not interchangeable for patients with keratoconus.
评估使用扫频源光相干断层扫描(Casia 2)和双 Scheimpflug(Galilei G6)断层摄影术测量圆锥角膜的眼生物测量学的一致性。
本回顾性研究纳入了 102 例圆锥角膜患者的 102 只眼,这些患者均使用这两种设备进行了检查。比较的参数包括平(Kf)和陡(Ks)角膜曲率计、前、后和总角膜曲率计的散光、中央(CCT)和最薄(TCT)角膜厚度。为了评估一致性,使用组内相关系数(ICC)和 Bland-Altman 分析,以及 95%的一致性界限(LoA)。
前向和总 Ks 和 Kf 具有中度或高度一致性,95%的一致性界限(LoA)范围超过 9.75 D。Galilei G6 中后向 Ks 和 Kf 较低(均 p<0.001)。前向、后向和总角膜曲率计的散光分别具有中度、较差和中度一致性。CCT 和 TCT 显示出极好的一致性;然而,95%的 LoA 范围超过 60μm。
对于诊断圆锥角膜和评估疾病进展的大多数参数,两种设备之间的一致性并不理想,差异具有临床意义。因此,对于圆锥角膜患者,这两种设备的测量结果不能互换。