Lasers Med Sci. 2024 Jul 31;39(1):201. doi: 10.1007/s10103-024-04155-3.
The evolution of laser medical devices for benign prostatic hyperplasia (BPH) treatment aims to enhance vaporization, coagulation, or tissue removal. In this study, we aim to evaluate the effectiveness and safety of the innovative application of insulated-gate bipolar transistor (IGBT) xenon lamp-pulsed drive technology holmium laser in endoscopic prostate enucleation operations using canine models. Six canines were used as an experimental unit, the breed of the canine unit used was beagle. Each canine served as its own control to minimize the number of experimental units. Endoscopic enucleation, performed by a single surgeon, involved enucleating the left hemi-prostate, leaving the right hem-prostate untouched to serve as the control. Throughout the study period, all canines maintained good health. No adverse events were observed in all six canines. Postoperatively, there were no indications of redness, swelling, or other adverse effects at the surgical sites. No abnormalities were observed in the appearance and morphology of major organs. The prostate and bladder, removed for further pathological evaluation, exhibited no abnormalities in size, color, or texture. No abnormalities or inflammation were observed, and the tissues were free of adhesions, indicating successful healing. In conclusion, our comparison of preoperative and postoperative parameters in canines suggests that the IGBT pulsed laser, at a power setting of 100 W, demonstrates characteristics of safety, efficacy, minimal tissue damage, and no major postoperative complications. This study establishes a theoretical foundation for future applications in human settings, encouraging further exploration of the IGBT holmium laser's potential in clinical practice.
激光医学设备在良性前列腺增生(BPH)治疗中的发展旨在增强汽化、凝固或组织去除效果。在本研究中,我们旨在评估使用绝缘栅双极晶体管(IGBT)氙灯脉冲驱动技术的创新钬激光在犬模型内窥镜前列腺剜除手术中的有效性和安全性。
使用 6 只犬作为实验单位,犬的品种为比格犬。每只犬作为自身对照,以最小化实验单位的数量。由一位外科医生进行内窥镜剜除术,包括剜除左侧前列腺,留下右侧前列腺作为对照。
在整个研究期间,所有犬都保持健康。在所有 6 只犬中都没有观察到不良事件。术后,手术部位无红肿或其他不良反应迹象。主要器官的外观和形态无异常。切除用于进一步病理评估的前列腺和膀胱,大小、颜色或质地无异常。未见异常或炎症,组织无粘连,表明愈合良好。
总之,我们对犬的术前和术后参数的比较表明,IGBT 脉冲激光在 100 W 的功率设置下具有安全、有效、最小组织损伤和无重大术后并发症的特点。本研究为未来在人类环境中的应用奠定了理论基础,鼓励进一步探索 IGBT 钬激光在临床实践中的潜力。