眼表分析仪LacryDiag的观察者内和观察者间可靠性

Intra- and interobserver reliability of ocular surface analyzer LacryDiag.

作者信息

Kos Eva, Cigić Valentina, Bušić Mladen, Bjeloš Mirjana, Miletić Daliborka, Kuzmanović Elabjer Biljana

机构信息

University Eye Clinic, University Hospital "Sveti Duh", Sveti Duh 64, Zagreb 10000, Croatia.

Faculty of Dental Medicine and Health Care Osijek, Faculty of Medicine Osijek, University Josip Juraj Strossmayer in Osijek, Osijek 31000, Croatia.

出版信息

Int J Ophthalmol. 2024 Aug 18;17(8):1418-1422. doi: 10.18240/ijo.2024.08.05. eCollection 2024.

Abstract

AIM

To invastigate intra- and interobserver reliability of interferometry, tear meniscus height (TMH) measurement and meibography (MBG) of an ocular surface analyzer, LacryDiag (Quantel Medical, France).

METHODS

Five consecutive measurements and subsequent analysis of interferometry, TMH, and MBG were recorded by two examiners using the LacryDiag. To assess intra- and interobserver reliability, we used Cohen's kappa for categorical variables (interferometry), or intraclass correlation coefficient for continuous variables (TMH, MBG).

RESULTS

Thirty eyes of 30 examinees were included. For both observers, there was excellent intraobserver reliability for MBG (0.955 and 0.970 for observer 1 and 2, respectively). Intraobserver reliability for observer 1 was substantial for interferometry (0.799), and excellent for TMH (0.863). Reliability for observer 2 was moderate for interferometry (0.535) and fair to good for TMH (0.431). Interobserver reliability was poor for interferometry (0.074) and fair to good for TMH (0.680) and MBG (0.414).

CONCLUSION

LacryDiag ocular surface analyzer in our study proves to be a reliable noninvasive tool for the evaluation of TMH and MBG. As for interferometry, poor interobserver reliability, fair to good intraobserver reliability for observer 1, and moderate for observer 2, leave room for improvement.

摘要

目的

研究法国Quantel Medical公司生产的眼表分析仪LacryDiag的干涉测量、泪膜半月板高度(TMH)测量及睑板腺照相(MBG)在观察者内和观察者间的可靠性。

方法

两名检查者使用LacryDiag对干涉测量、TMH和MBG进行连续五次测量并随后进行分析。为评估观察者内和观察者间的可靠性,对于分类变量(干涉测量)我们使用Cohen's kappa系数,对于连续变量(TMH、MBG)我们使用组内相关系数。

结果

纳入30名受试者的30只眼。对于两名观察者,MBG的观察者内可靠性均极佳(观察者1和2分别为0.955和0.970)。观察者1的干涉测量观察者内可靠性为实质性(0.799),TMH为极佳(0.863)。观察者2的干涉测量可靠性为中等(0.535),TMH为一般至良好(0.431)。观察者间干涉测量可靠性较差(0.074),TMH为一般至良好(0.680),MBG为中等(0.414)。

结论

在我们的研究中,LacryDiag眼表分析仪被证明是评估TMH和MBG的可靠无创工具。至于干涉测量,观察者间可靠性较差,观察者1的观察者内可靠性一般至良好,观察者2的为中等,仍有改进空间。

相似文献

1
Intra- and interobserver reliability of ocular surface analyzer LacryDiag.
Int J Ophthalmol. 2024 Aug 18;17(8):1418-1422. doi: 10.18240/ijo.2024.08.05. eCollection 2024.
3
Reliability and clinical applicability of a novel tear film imaging tool.
Graefes Arch Clin Exp Ophthalmol. 2021 Jul;259(7):1935-1943. doi: 10.1007/s00417-021-05162-8. Epub 2021 Mar 29.
4
Reliability, repeatability, and accordance between three different corneal diagnostic imaging devices for evaluating the ocular surface.
Front Med (Lausanne). 2022 Jul 29;9:893688. doi: 10.3389/fmed.2022.893688. eCollection 2022.
5
Reliability analysis for digital adolescent idiopathic scoliosis measurements.
J Spinal Disord Tech. 2005 Apr;18(2):152-9. doi: 10.1097/01.bsd.0000148094.75219.b0.
8
Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.
J Cataract Refract Surg. 2001 Dec;27(12):1961-8. doi: 10.1016/s0886-3350(01)01214-7.
9
Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.
Graefes Arch Clin Exp Ophthalmol. 2011 Jan;249(1):83-7. doi: 10.1007/s00417-010-1546-z. Epub 2010 Oct 28.
10
Interobserver variability of an open-source software for tear meniscus height measurement.
Cont Lens Anterior Eye. 2016 Aug;39(4):249-56. doi: 10.1016/j.clae.2016.02.009. Epub 2016 Mar 3.

本文引用的文献

1
Dynamic tear meniscus parameters in complete blinking: insights into dry eye assessment.
Int J Ophthalmol. 2023 Dec 18;16(12):1911-1918. doi: 10.18240/ijo.2023.12.01. eCollection 2023.
2
Evaluation of ocular surface and tear function - A review of current approaches for dry eye.
Indian J Ophthalmol. 2022 Jun;70(6):1883-1891. doi: 10.4103/ijo.IJO_1804_21.
3
Advances in Dry Eye Disease Examination Techniques.
Front Med (Lausanne). 2022 Jan 25;8:826530. doi: 10.3389/fmed.2021.826530. eCollection 2021.
4
Efficacy of an Intranasal Tear Neurostimulator in Sjögren Syndrome Patients.
Clin Ophthalmol. 2021 Oct 27;15:4291-4296. doi: 10.2147/OPTH.S312108. eCollection 2021.
5
Quantitative analysis of morphological and functional features in Meibography for Meibomian Gland Dysfunction: Diagnosis and Grading.
EClinicalMedicine. 2021 Sep 11;40:101132. doi: 10.1016/j.eclinm.2021.101132. eCollection 2021 Oct.
7
Reliability and clinical applicability of a novel tear film imaging tool.
Graefes Arch Clin Exp Ophthalmol. 2021 Jul;259(7):1935-1943. doi: 10.1007/s00417-021-05162-8. Epub 2021 Mar 29.
8
Evaluation of tear meniscus height using different clinical methods.
Clin Exp Optom. 2021 Jul;104(5):583-588. doi: 10.1080/08164622.2021.1878854. Epub 2021 Mar 2.
9
Comparison of meibomian gland loss area measurements between two computer programs and intra-inter-observer agreement.
Int Ophthalmol. 2020 May;40(5):1261-1267. doi: 10.1007/s10792-020-01292-w. Epub 2020 Jan 23.
10
Repeatability of Noninvasive Keratograph 5M Measurements Associated With Contact Lens Wear.
Eye Contact Lens. 2019 Nov;45(6):377-381. doi: 10.1097/ICL.0000000000000596.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验