Suppr超能文献

泪液诊断眼表面分析仪评估睑板腺和泪膜重复性的评估。

Evaluation of the Repeatability of the LacryDiag Ocular Surface Analyzer for Assessment of the Meibomian Glands and Tear Film.

机构信息

Department of Ophthalmology, The University of Texas Southwestern Medical Center, Dallas, TX, USA.

出版信息

Transl Vis Sci Technol. 2021 Aug 2;10(9):1. doi: 10.1167/tvst.10.9.1.

Abstract

PURPOSE

This study aimed to determine the intra- and interobserver repeatability of the new LacryDiag Ocular Surface Analyzer and compare it to a similar all-in-one device, the OCULUS Keratograph 5M.

METHODS

Thirty healthy subjects aged 18 years and above were recruited for this study. All patients were free of any existing ocular pathology. The LacryDiag Ocular Surface Analyzer was used to evaluate tear meniscus height, interferometry, noninvasive tear break-up time (NIBUT), and meibography. The same or analogous exams were performed using the OCULUS Keratograph 5M. Test equivalation was used to compare data from corresponding examinations. Paired t-tests and coefficient of variation were used to determine inter- and intraobserver repeatability. Bland-Altman analysis was used to determine level of agreement between devices.

RESULTS

There were no differences in mean values for tear meniscus height, NIBUT, or tear film interferometry between observers for either device. Significant differences were found between observers for meibography when using the LacryDiag (P = 0.008 for percent loss calculation and P = 0.004 for grading scale). Intra-observer variability for NIBUT was significantly higher for the Keratograph (P = 0.0003 for observer A and P < 0.0001 for observer B).

CONCLUSIONS

There was a good correlation but poor agreement between devices for a given observer. This was likely influenced by the use of repeated testing and the non-dry eye cohort.

TRANSLATIONAL RELEVANCE

Both the repeatability of the testing device and the use of multiple outcome measures are essential for the diagnosis and monitoring of patients with dry eye disease (DED).

摘要

目的

本研究旨在确定新型 LacryDiag 眼表面分析仪的观察者内和观察者间重复性,并将其与类似的一体式设备 Oculu 角膜地形图仪 5M 进行比较。

方法

本研究招募了 30 名年龄在 18 岁及以上的健康受试者。所有患者均无任何现患眼疾。使用 LacryDiag 眼表面分析仪评估泪膜弯月高、干涉测量、非侵入性泪膜破裂时间(NIBUT)和睑板腺照相。使用 Oculu 角膜地形图仪进行相同或类似的检查。采用等效性检验比较相应检查的结果。采用配对 t 检验和变异系数来确定观察者内和观察者间的重复性。采用 Bland-Altman 分析来确定设备间的一致性水平。

结果

对于两种设备,观察者间的泪膜弯月高、NIBUT 或泪膜干涉测量值均值无差异。当使用 LacryDiag 时,观察者间的睑板腺照相结果存在显著差异(计算百分比损失时 P = 0.008,分级量表时 P = 0.004)。对于 Keratograph,NIBUT 的观察者内变异性显著更高(观察者 A 时 P = 0.0003,观察者 B 时 P < 0.0001)。

结论

对于特定观察者,两种设备之间相关性良好,但一致性较差。这可能是由于重复测试和非干眼队列的使用所导致的。

翻译

温伟强

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/995d/8340660/50de641da4b7/tvst-10-9-1-f001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验