Suppr超能文献

载脂蛋白A-II的比浊法测定

Turbidimetric assay of apolipoprotein A-II.

作者信息

Schriewer H, Emke F, Assmann G

出版信息

J Clin Chem Clin Biochem. 1985 Jun;23(6):355-9. doi: 10.1515/cclm.1985.23.6.355.

Abstract

A turbidimetric assay for the quantification of apolipoprotein A-II in serum is described. Its adaptation for the Cobas Bio Analyser (Hoffmann La Roche) is reported. Regression analysis of apolipoprotein A-II values measured by turbidimetry revealed a good measure of agreement with the data obtained by radial immunodiffusion (RID) (r = 0.90, y = 0.97x + 0.021, n = 100) as well as with data obtained by immunonephelometry (r = 0.838, y = 0.925x + 0.029, n = 34). A variation coefficient of 3.2% (n = 18) was found in relation to the precision in the series and a variation coefficient of 3.5% (n = 36) in relation to day to day precision.

摘要

本文描述了一种用于定量测定血清中载脂蛋白A-II的比浊法。报告了其在Cobas Bio分析仪(霍夫曼·罗氏公司)上的应用。通过比浊法测定的载脂蛋白A-II值的回归分析表明,与通过放射免疫扩散法(RID)获得的数据(r = 0.90,y = 0.97x + 0.021,n = 100)以及与通过免疫比浊法获得的数据(r = 0.838,y = 0.925x + 0.029,n = 34)具有良好的一致性。相对于系列中的精密度,发现变异系数为3.2%(n = 18),相对于每日精密度,变异系数为3.5%(n = 36)。