Suppr超能文献

跨平台差异对经胸超声心动图测量及临床诊断的影响。

Effect of cross-platform variations on transthoracic echocardiography measurements and clinical diagnosis.

作者信息

Hashemi Mohammad Saber, Farsiani Yasaman, Pressman Gregg S, Amini M Reza, Kheradvar Arash

机构信息

Department of Biomedical Engineering, University of California Irvine, Irvine, CA, USA.

Division of Cardiology, Thomas Jefferson University, Thomas Jefferson Einstein Hospital, Philadelphia, PA, USA.

出版信息

Eur Heart J Imaging Methods Pract. 2024 Sep 23;2(3):qyae097. doi: 10.1093/ehjimp/qyae097. eCollection 2024 Jul.

Abstract

AIMS

Accurate cardiac chamber quantification is essential for clinical decisions and ideally should be consistent across different echocardiography systems. This study evaluates variations between the Philips EPIQ CVx (version 9.0.3) and Canon Aplio i900 (version 7.0) in measuring cardiac volumes, ventricular function, and valve structures.

METHODS AND RESULTS

In this gender-balanced, single-centre study, 40 healthy volunteers (20 females and 20 males) aged 40 years and older (mean age 56.75 ± 11.57 years) were scanned alternately with both systems by the same sonographer using identical settings for both 2D and 4D acquisitions. We compared left ventricular (LV) and right ventricular (RV) volumes using paired -tests, with significance set at < 0.05. Correlation and Bland-Altman plots were used for quantities showing significant differences. Two board-certified cardiologists evaluated valve anatomy for each platform. The results showed no significant differences in LV end-systolic volume and LV ejection fraction between platforms. However, LV end-diastolic volume (LVEDV) differed significantly (biplane: = 0.018; 4D: = 0.028). Right ventricular (RV) measurements in 4D showed no significant differences, but there were notable disparities in 2D and 4D volumes within each platform ( < 0.01). Significant differences were also found in the LV systolic dyssynchrony index ( = 0.03), LV longitudinal strain ( = 0.04), LV twist ( = 0.004), and LV torsion ( = 0.005). Valve structure assessments varied, with more abnormalities noted on the Philips platform.

CONCLUSION

Although LV and RV volumetric measurements are generally comparable, significant differences in LVEDV, LV strain metrics, and 2D vs. 4D measurements exist. These variations should be considered when using different platforms for patient follow-ups.

摘要

目的

准确的心脏腔室定量对于临床决策至关重要,理想情况下,不同超声心动图系统的测量结果应保持一致。本研究评估了飞利浦EPIQ CVx(9.0.3版本)和佳能Aplio i900(7.0版本)在测量心脏容积、心室功能和瓣膜结构方面的差异。

方法与结果

在这项性别均衡的单中心研究中,40名年龄40岁及以上(平均年龄56.75±11.57岁)的健康志愿者(20名女性和20名男性)由同一名超声检查人员使用相同的二维和四维采集设置,交替使用这两种系统进行扫描。我们使用配对t检验比较左心室(LV)和右心室(RV)容积,显著性设定为P<0.05。对于显示出显著差异的量,使用相关性分析和布兰德-奥特曼图。两名经过董事会认证的心脏病专家对每个平台的瓣膜解剖结构进行评估。结果显示,各平台之间左心室收缩末期容积和左心室射血分数无显著差异。然而,左心室舒张末期容积(LVEDV)存在显著差异(双平面:P = 0.018;四维:P = 0.028)。四维测量中的右心室(RV)测量结果无显著差异,但每个平台内的二维和四维容积存在显著差异(P<0.01)。在左心室收缩不同步指数(P = 0.03)、左心室纵向应变(P = 0.04)、左心室扭转(P = 0.004)和左心室扭转角度(P = 0.005)方面也发现了显著差异。瓣膜结构评估结果存在差异,飞利浦平台上发现的异常更多。

结论

尽管左心室和右心室容积测量结果总体上具有可比性,但左心室舒张末期容积、左心室应变指标以及二维与四维测量结果存在显著差异。在使用不同平台进行患者随访时,应考虑这些差异。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/dd0e/11465166/0c5c7ba7d73e/qyae097_ga.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验