Accuracy and comparability of long-term measurements of cholesterol.

作者信息

Hainline A, Karon J M, Winn C L, Gill J B

出版信息

Clin Chem. 1986 Apr;32(4):611-5.

PMID:3955810
Abstract

Laboratories of the Lipid Research Clinics Program (LRC) maintained the accuracy of their measurements of total cholesterol by using seven pooled serum calibrators supplied by the Centers for Disease Control (CDC). Over the 11-year life of the LRC, each calibrator was prepared in succession and a target value was assigned by the CDC reference method for cholesterol. The results of a special experiment in which six of the seven calibrators were analyzed simultaneously demonstrated that the target values were accurately assigned. Deviations of the target values from the experimental means ranged from zero to 1.7% of the original target value. The experiment revealed no evidence of drift in the bias of the reference method over the life of LRC and demonstrated the accuracy, consistency, and the comparability of the values assigned to the successive calibrator pools used by the LRC laboratories during more than eight years. It demonstrated the reliability of a reference method and the suitability of frozen serum pools for maintaining an accurate measurement base for serum cholesterol.

摘要

脂质研究临床项目(LRC)的实验室通过使用疾病控制中心(CDC)提供的七种混合血清校准物来保持其总胆固醇测量的准确性。在LRC的11年存续期内,每种校准物依次制备,且疾病控制中心的胆固醇参考方法为其指定了一个目标值。一项特殊实验的结果表明,目标值被准确指定,在该实验中,七个校准物中的六个被同时分析。目标值与实验平均值的偏差范围为零至原始目标值的1.7%。该实验未发现疾病控制中心参考方法的偏差在LRC存续期内存在漂移的证据,并证明了在超过八年的时间里,脂质研究临床项目实验室使用的连续校准物池所分配值的准确性、一致性和可比性。它证明了参考方法的可靠性以及冷冻血清池对于维持血清胆固醇准确测量基础的适用性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验