Matrix effects on proficiency testing materials. Impact on accuracy of cholesterol measurement in laboratories in the nation's largest hospital system.

作者信息

Naito H K, Kwak Y S, Hartfiel J L, Park J K, Travers E M, Myers G L, Ross J W, Eckfeldt J H, Hartmann A E

机构信息

Department of Veterans Affairs, National Center for Laboratory Accuracy and Standardization, Veterans Affairs Medical Center, Cleveland, Ohio.

出版信息

Arch Pathol Lab Med. 1993 Apr;117(4):345-51.

PMID:8466396
Abstract

The objective of this collaborative study with the Department of Veterans Affairs (VA), College of American Pathologists (CAP), and the Centers for Disease Control and Prevention (CDC) was to quantitate the matrix-induced biases of cholesterol measurements on the CAP Comprehensive Chemistry Surveys materials used in proficiency testing (PT). A total of 174 VA Medical Centers outpatient clinics and clinical laboratories participate in the VA-CDC National Cholesterol Standardization and Certification Program. This study was conducted in 112 VA laboratories that have been standardized for measuring cholesterol accurately (within +/- 3.0% of the CDC reference-method values) using fresh, unfrozen, unadulterated human serum samples. Fresh serum samples and 1990 CAP Surveys materials were sent by overnight mail, and the laboratories were asked to analyze them simultaneously in triplicate in a single analytic batch run. The results showed significant matrix-effect biases with the CAP Surveys materials with six of the eight major peer groups, despite the fact that accuracy of cholesterol measurements was maintained with fresh serum samples. The magnitude and direction (positive or negative) of the matrix-effect biases were instrument, reagent, and method specific using the following peer groups: du Pont Dimension (-8.9%); Beckman CX4, CX5, and CX7 (-5.5%); Kodak Ektachem 400, 500, and 700 (+4.4%); Instrumentation Laboratory Monarch (-3.1%); Baxter Paramax (-2.4%); Technicon SMAC and RA (+1.3%); Hitachi/BMD 704 through 747 (+0.4%); and Abbott Spectrum (-0.3%). The CAP PT materials used currently do not behave in a manner identical to fresh human serum when measuring cholesterol on many, but not all, analytic systems. The observed biases due to "matrix effects" with PT materials will cause incorrect conclusions about the accuracy of many laboratory procedures performed on fresh patient specimens. This matrix-effect phenomenon will severely hamper interlaboratory accuracy transfer, standardization efforts, and monitoring performance of a laboratory's testing accuracy with the use of the current survey materials used in PT programs. Collaborative efforts are needed to (1) improve PT fluids to analytically behave more like fresh, human serum; (2) improve instrument design and reagent formulation; and (3) select methods and methodologic parameters that are more "robust" and less sensitive to the exact character of processed calibrators, quality control, and PT materials.

摘要

这项与美国退伍军人事务部(VA)、美国病理学家学会(CAP)以及疾病控制与预防中心(CDC)合作开展的研究,旨在对能力验证(PT)中使用的CAP综合化学检测材料上胆固醇测量的基质诱导偏差进行定量分析。共有174家VA医疗中心的门诊诊所和临床实验室参与了VA - CDC国家胆固醇标准化与认证项目。本研究在112家已通过使用新鲜、未冷冻、未掺假的人血清样本准确测量胆固醇(在CDC参考方法值的±3.0%范围内)而实现标准化的VA实验室中进行。新鲜血清样本和1990年的CAP检测材料通过隔夜快递发送,要求实验室在单个分析批次中对其进行一式三份的同步分析。结果显示,尽管使用新鲜血清样本时胆固醇测量的准确性得以维持,但在八个主要同类组中的六个组中,CAP检测材料存在显著的基质效应偏差。使用以下同类组时,基质效应偏差的大小和方向(正或负)因仪器、试剂和方法而异:杜邦Dimension(-8.9%);贝克曼CX4、CX5和CX7(-5.5%);柯达Ektachem 400、500和700(+4.4%);仪器实验室Monarch(-3.1%);百特Paramax(-2.4%);Technicon SMAC和RA(+1.3%);日立/BMD 704至747(+0.4%);以及雅培Spectrum(-0.3%)。在许多(但并非所有)分析系统上测量胆固醇时,当前使用的CAP PT材料的表现与新鲜人血清并不相同。PT材料因“基质效应”而观察到的偏差,将导致对许多针对新鲜患者样本所执行实验室程序的准确性得出错误结论。这种基质效应现象将严重阻碍实验室间准确性传递、标准化工作以及使用PT项目中当前检测材料对实验室检测准确性的监测。需要共同努力来:(1)改进PT液体,使其在分析性能上更接近新鲜人血清;(2)改进仪器设计和试剂配方;(3)选择对处理后的校准品、质量控制品和PT材料的精确特性更“稳健”且更不敏感的方法和方法学参数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验