三种 Scheimpflug 断层扫描仪和一种 Placido 角膜地形图仪测量前节的比较。

Comparison of anterior segment measurements by 3 Scheimpflug tomographers and 1 Placido corneal topographer.

机构信息

G.B. Bietti Eye Foundation-IRCCS, Rome, Italy.

出版信息

J Cataract Refract Surg. 2011 Sep;37(9):1679-85. doi: 10.1016/j.jcrs.2011.03.055.

Abstract

PURPOSE

To compare the anterior segment measurements provided by 3 Scheimpflug tomographers and a Placido corneal topographer.

SETTING

Private clinical ophthalmology practice.

DESIGN

Evaluation of diagnostic test or technology.

METHODS

In a sample of 25 consecutive patients having either refractive or cataract surgery, the anterior eye segment was analyzed by means of a rotating Scheimpflug camera (Pentacam), 2 devices with a Scheimpflug camera combined with a Placido disk (Sirius and TMS-5), and a Placido disk corneal topographer (Keratron). Measurement results were compared using analysis of variance. Agreement was assessed using Bland-Altman plots.

RESULTS

The mean simulated keratometry (K) was different between the 4 instruments (P<.0001), with Keratron providing the highest value (44.43 diopters [D] ± 1.28 [SD]). The Pentacam and Sirius provided the lowest values (44.05 ± 1.21 D and 44.05 ± 1.27 D, respectively), without statistical difference (posttest). The mean posterior corneal power and minimum corneal thickness were statistically different between the 3 Scheimpflug cameras (P<.0001 and P=.0210, respectively); 95% limits of agreement, however, were narrow for posterior corneal power and large for corneal thickness. The only 2 devices measuring the distance between the corneal endothelium and the anterior lens surface showed a statistically but not clinically significant difference (2.90 ± 0.48 mm and 2.94 ± 0.47 mm, respectively). There were no statistically significant differences in anterior corneal asphericity between the 4 instruments.

CONCLUSION

Although the measurements of some parameters by different instruments were similar, caution is warranted before using them interchangeably.

摘要

目的

比较 3 种 Scheimpflug 断层扫描仪和 Placido 角膜地形图仪提供的眼前节测量值。

设置

私人临床眼科诊所。

设计

诊断测试或技术的评估。

方法

在 25 例连续接受屈光或白内障手术的患者中,使用旋转 Scheimpflug 相机(Pentacam)、2 种结合 Scheimpflug 相机和 Placido 盘的设备(Sirius 和 TMS-5)以及 Placido 盘角膜地形图仪(Keratron)对眼前节进行分析。使用方差分析比较测量结果。使用 Bland-Altman 图评估一致性。

结果

4 种仪器的模拟角膜曲率(K)平均值不同(P<.0001),Keratron 提供的数值最高(44.43 屈光度 [D] ± 1.28 [SD])。Pentacam 和 Sirius 提供的数值最低(分别为 44.05 ± 1.21 D 和 44.05 ± 1.27 D,无统计学差异)。3 种 Scheimpflug 相机的后角膜曲率和最小角膜厚度平均值存在统计学差异(P<.0001 和 P=.0210);然而,95%的一致性界限在后角膜曲率方面较窄,在角膜厚度方面较大。仅 2 种测量角膜内皮和前晶状体表面之间距离的设备显示出统计学但无临床意义的差异(分别为 2.90 ± 0.48 毫米和 2.94 ± 0.47 毫米)。4 种仪器的前角膜非球面性无统计学差异。

结论

尽管不同仪器的某些参数测量值相似,但在相互替换使用之前应谨慎。