种植体的外六边和锥形连接体的共振频率分析受厌氧胶影响的体外研究。

Influence of anaerobic adhesive on the resonance frequency analysis measured in abutments with external hexagon and conical connections: an in vitro study.

机构信息

Department of Restorative Dentistry, School of Dentistry, Universidade Federal de Minas Gerais, Avenida Antônio Carlos - 6627, Pampulha, Belo Horizonte, MG, Brazil.

出版信息

Clin Oral Investig. 2024 Nov 29;28(12):667. doi: 10.1007/s00784-024-06057-8.

Abstract

OBJECTIVES

To evaluate whether the application of an anaerobic adhesive (AA) to the abutment screws interferes with the implant stability quotient (ISQ) values in abutments with external hexagon (EHC) and conical connections (CC).

MATERIALS AND METHODS

20 implants 4.0 mm x 10 mm were used, with similar formats, 10 with EHC and 10 with CC, inserted by the same operator into fresh bovine ribs. The abutment was then screwed with a high precision digital torque meter, and resonance frequency was measured. The abutments were then unscrewed, a drop of Loctite 242 adhesive was applied along the entire length of the thread, the abutments were screwed again (Groups EHC-AA and CC-AA). The ISQ was registered immediately, 24 h, and 72 h after installation. Data were analyzed using descriptive statistics and comparison tests (p < 0.05).

RESULTS

ISQ values measured in abutments without AA were significantly higher than those measured after applying the adhesive, for EHC and CC. The one-way repeated measures ANOVA showed that, in the intragroup analysis, there was no statistically significant difference in the ISQ, comparing the three measurement moments after applying the adhesive, irrespective of connection system (CC: F = 0.191, p = 0.827; EHC: F = 0.323, p = 727), which suggests that the adhesive curing does not affect the ISQ.

CONCLUSIONS

The use of an AA on abutment screws had a significant effect on RFA measurements, for EHC and CC. The curing time of the AA did not affect the ISQ values measured in the abutments.

CLINICAL RELEVANCE

The ISQ values cannot be compared in situations with and without AA.

摘要

目的

评估在具有外六方(EHC)和锥形连接(CC)的基台上应用厌氧胶(AA)是否会干扰基台螺钉的种植体稳定性系数(ISQ)值。

材料与方法

使用 20 个 4.0mm x 10mm 的种植体,具有相似的格式,其中 10 个具有 EHC,10 个具有 CC,由同一位操作员插入新鲜牛肋骨中。然后,使用高精度数字扭矩计拧紧基台,并测量共振频率。然后将基台拧下,在整个螺纹上涂抹一滴 Loctite 242 胶,再次拧紧基台(EHC-AA 和 CC-AA 组)。安装后立即、24 小时和 72 小时记录 ISQ 值。使用描述性统计和比较检验(p<0.05)对数据进行分析。

结果

未使用 AA 的基台测量的 ISQ 值明显高于应用 AA 后测量的值,对于 EHC 和 CC 基台都是如此。单向重复测量方差分析显示,在组内分析中,无论连接系统如何(CC:F=0.191,p=0.827;EHC:F=0.323,p=727),在应用 AA 后三个测量时刻的 ISQ 均无统计学差异,这表明 AA 固化不会影响 ISQ。

结论

在 EHC 和 CC 基台上使用 AA 对 RFA 测量有显著影响。AA 的固化时间不会影响基台上测量的 ISQ 值。

临床相关性

在有和没有 AA 的情况下,ISQ 值不能进行比较。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验