一种新型工具的开发与验证,该工具用于以每分钟体积单位(微升)测量泪液量,即“标准国际单位(SI)泪液试纸” 。

Development and validation of a novel tool to measure tear volume in volumetric unit (microlitre) per minute "the standard international unit (SI) tear strip".

作者信息

Balamurugan R, Bhattacharyya Anusuya, Sarma Phulen

机构信息

Department of Ophthalmology, All India Institute of Medical Sciences, Mangalagiri, Andhra Pradesh, India.

Department of Ophthalmology, All India Institute of Medical Sciences, Guwahati, Assam, India.

出版信息

Indian J Ophthalmol. 2025 Jan 1;73(1):141-145. doi: 10.4103/IJO.IJO_162_24. Epub 2024 Dec 23.

Abstract

The Schirmer strip measures tear volume in millimeters. A major limitation of the Schirmer strip is that it does not measure tear volume in a volumetric unit. Furthermore, the linearity range, lower limit of detection (LLOD), and lower limit of quantification (LLOQ) are not established. In this context, we developed a new filter paper based tear volume strip, which measures tear volume in microlitres (a volumetric unit), with a defined linearity range, LLOD, LLOQ, as well as inter-day and intra-day variability. We used standard Whatman filter paper no. 41, measuring 40 × 5 mm. Balanced salt solution (BSS) was used as a tear surrogate. By placing a specific volume of BSS at one marked end of the paper, we measured the distance travelled by the BSS after one minute. Different volumes of BSS (0.5, 1, 2, 4, 8, and 16 μl) were used, and the distance travelled by the BSS drop was recorded. Linearity was evaluated, and the LLOQ and LLOD were determined. Scale data were generated within the linearity range (10 replicates for each). Percentage recovery was calculated, and inter-day and intra-day variation as well as intra examiner and inter examiner coefficients of variation (CV%) were evaluated. High linearity (r2 = 0.976) was observed across all the tested volumes (0.5-16 μl), However, linearity improved within the tested volume range of 1-8 μl (r2 = 0.996). With an LLOD of 0.538 μl and an LLOQ of 1.63 μl, data for the scale were generated from 2-8 μl. Overall, volumetric recovery ranged from 97.34%-104%. The intra-day CV% ranged from 4.576% (for 2 μl) to 1.911% (for 5 μl) and 1.71% (for 7 μl). The inter-day CV% ranged from 1.715 (7 μl) to 4.36% (for 2 μl). The intra examiner CV% ranged from 0.422% (for 7 μl) to 2.66% (for 2 μl). The inter examiner CV% ranged from 1.93% (7 μl) to 3.69% (for 2 μl). This new tear volume measuring tool (named SI tear strip) represents a valuable tool for the measurement of tear volume (in μl) per unit time. This tool may be very helpful for the quantitative evaluation of dry eye patients and functional assessment of lacrimal glands.

摘要

施密特试纸以毫米为单位测量泪液量。施密特试纸的一个主要局限性在于它不以容积单位来测量泪液量。此外,其线性范围、检测下限(LLOD)和定量下限(LLOQ)尚未确定。在此背景下,我们开发了一种基于滤纸的新型泪液量试纸,它以微升(容积单位)来测量泪液量,具有明确的线性范围、LLOD、LLOQ以及日间和日内变异性。我们使用了标准的Whatman 41号滤纸,尺寸为40×5毫米。平衡盐溶液(BSS)用作泪液替代物。通过在试纸的一个标记端放置特定体积的BSS,我们测量了一分钟后BSS移动的距离。使用了不同体积的BSS(0.5、1、2、4、8和16微升),并记录了BSS液滴移动的距离。评估了线性度,并确定了LLOQ和LLOD。在整个线性范围内(每个体积进行10次重复)生成了刻度数据。计算了回收率,并评估了日间和日内变化以及检查者内和检查者间变异系数(CV%)。在所有测试体积(0.5 - 16微升)中均观察到高线性度(r2 = 0.976),然而,在1 - 8微升的测试体积范围内线性度有所提高(r2 = 0.996)。LLOD为0.538微升,LLOQ为1.63微升,刻度数据是从2 - 8微升生成的。总体而言,容积回收率在97.34% - 104%之间。日内CV%范围从4.576%(对于2微升)到1.911%(对于5微升)和1.71%(对于7微升)。日间CV%范围从1.715(对于7微升)到4.36%(对于2微升)。检查者内CV%范围从0.422%(对于7微升)到2.66%(对于2微升)。检查者间CV%范围从1.93%(对于7微升)到3.69%(对于2微升)。这种新型泪液量测量工具(命名为SI泪液试纸)是一种用于测量单位时间内泪液量(以微升为单位)的有价值工具。该工具对于干眼患者的定量评估和泪腺的功能评估可能非常有帮助。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1890/11831932/187e286e0367/IJO-73-141-g002.jpg