Buchanan R A, Williams T D
Am J Optom Physiol Opt. 1985 Jan;62(1):59-62. doi: 10.1097/00006324-198501000-00008.
Intraocular pressures (IOP's) were measured using the Digilab Pneuma-tonometer with the subject in both the sitting and supine positions. The IOP with the Pneuma-tonometer was greater (17.03 mm Hg) in the supine position than in the sitting position (12.90 mm Hg). The IOP measured with the Pneuma-tonometer, with the subject sitting, was similar to the IOP measured with the Goldmann applanation tonometer (13.42 mm Hg). Inasmuch as IOP tends to be lower with the patient in a sitting position, the clinician should be alert to the possibility that some patients with borderline Goldmann IOP's may have pressures well above the normal range when they are lying down. Due to the continuous recording provided by the Pneuma-tonometer, the ocular pulse pressure can be measured. We found a mean ocular pulse pressure amplitude of 1.77 mm Hg. All values are shown plus or minus 1 SD. Comparison of the ocular pulse pressure amplitude for the two eyes could aid in detecting patients with suspected carotid artery stenosis.
使用Digilab气动眼压计在受试者坐位和仰卧位时测量眼压。气动眼压计测量的眼压在仰卧位(17.03毫米汞柱)高于坐位(12.90毫米汞柱)。受试者坐位时用气动眼压计测量的眼压与用戈德曼压平眼压计测量的眼压(13.42毫米汞柱)相似。由于患者坐位时眼压往往较低,临床医生应警惕一些戈德曼眼压临界值的患者在躺下时眼压可能远高于正常范围的可能性。由于气动眼压计提供连续记录,因此可以测量眼动脉压。我们发现平均眼动脉压振幅为1.77毫米汞柱。所有值均显示为加或减1个标准差。比较双眼的眼动脉压振幅有助于检测疑似颈动脉狭窄的患者。