Hughes J G, Snyder R J, Washington J A
Diagn Microbiol Infect Dis. 1985 Mar;3(2):139-42. doi: 10.1016/0732-8893(85)90023-9.
A leukocyte esterase/nitrite (Chemstrip LNTM) and a bioluminescence assay (LumacTM) were evaluated for detection of bacteriuria in 1,000 urine specimens. Both devices provided high predictive negative values (95.4-97.8%); however, false-negative values at levels of bacteriuria greater than or equal to 10(4) CFU/ml were 22.6% and 12.3%, respectively, for the leukocyte esterase/nitrate and the bioluminescence assay tests. The corresponding false-negative values at levels of bacteriuria of greater than or equal to 10(5) CFU/ml were 16.5% and 3.9%, respectively.
对1000份尿液标本,评估了白细胞酯酶/亚硝酸盐(Chemstrip LNTM)和生物发光检测法(LumacTM)用于检测菌尿症的情况。两种检测设备均具有较高的预测阴性值(95.4 - 97.8%);然而,对于白细胞酯酶/硝酸盐检测和生物发光检测,当菌尿水平大于或等于10(4) CFU/ml时,假阴性值分别为22.6%和12.3%。当菌尿水平大于或等于10(5) CFU/ml时,相应的假阴性值分别为16.5%和3.9%。