Suppr超能文献

无法在自动微生物系统革兰氏阳性和革兰氏阴性药敏卡上充分控制抗菌药物。

Inability to adequately control antimicrobial agents on AutoMicrobic System Gram-Positive and Gram-Negative Susceptibility Cards.

作者信息

Kellogg J A

出版信息

J Clin Microbiol. 1985 Mar;21(3):454-6. doi: 10.1128/jcm.21.3.454-456.1985.

Abstract

The Vitek Gram-Positive Susceptibility Card (GPS) and Gram-Negative General Susceptibility Plus Card (GSC Plus) were tested on the AutoMicrobic System (AMS) 50 times each with the recommended control organisms. Only 1 drug (chloramphenicol) of 11 on the GPS and 1 (gentamicin) of 10 on the GSC Plus could be adequately controlled, leaving unsubstantiated the results obtained with patient isolates on the remaining 19 antimicrobial agents.

摘要

Vitek革兰氏阳性药敏卡(GPS)和革兰氏阴性通用药敏增强卡(GSC Plus)分别在自动微生物系统(AMS)上,使用推荐的对照微生物各测试了50次。GPS上的11种药物中只有1种(氯霉素)以及GSC Plus上的10种药物中只有1种(庆大霉素)能够得到充分控制,其余19种抗菌药物用于患者分离株所获得的结果无法得到证实。

相似文献

6
Clinical evaluation of new AMS GSC plus card.新型AMS GSC plus卡的临床评估