Study on the stability and accuracy of the new Booster portable cardiopulmonary function meter.

作者信息

Yun Hezhang, Zhang Wenbo, Yu Chen, Li Qiang, Song Yafeng

机构信息

China Institute of Sport and Health Science, Beijing Sport University, Beijing, China.

School of Physical Education, Zhejiang Guangsha Vocational and Technical University of Construction, Dongyang, China.

出版信息

Front Physiol. 2025 Jan 8;15:1453942. doi: 10.3389/fphys.2024.1453942. eCollection 2024.

Abstract

This study aims to assess the reliability and accuracy of a novel portable cardiopulmonary function meter, "Booster," developed by our research group, across various exercise intensities and modalities. The study was segmented into reliability and validity assessments. Twenty-two male participants underwent reliability testing, conducting two sequential tests on a treadmill while wearing the Booster to measure VO and VE among other parameters at increasing intensities. For validity testing, 64 participants were randomly divided into treadmill and cycle ergometer groups, with tests conducted using both the Booster and the Cortex Metalyzer 3B systems. Overall, the Booster demonstrated high retest reliability for VO and VE measurements during treadmill exercises, albeit showing poor consistency during rest and low-intensity exercise phases. Validity testing indicated no significant differences in VO and VE measurements between Booster and Cortex Metalyzer 3B across all exercise stages on both treadmill and cycle ergometer, suggesting good correlation. However, discrepancies in measurements between Booster and Cortex Metalyzer 3B were observed during rest and maximal exertion phases. The Booster exhibits commendable reliability and stability during most treadmill exercise phases and shows generally acceptable validity compared to the Cortex Metalyzer 3B system. Nonetheless, potential measurement discrepancies may occur during rest and maximal exertion conditions.

摘要

本研究旨在评估我们研究小组开发的新型便携式心肺功能测量仪“Booster”在各种运动强度和运动方式下的可靠性和准确性。该研究分为可靠性和有效性评估。22名男性参与者进行了可靠性测试,他们在跑步机上佩戴Booster进行了两次连续测试,以在强度增加时测量包括VO和VE在内的其他参数。对于有效性测试,64名参与者被随机分为跑步机组和自行车测力计组,使用Booster和Cortex Metalyzer 3B系统进行测试。总体而言,Booster在跑步机运动期间对VO和VE测量显示出较高的重测可靠性,尽管在休息和低强度运动阶段一致性较差。有效性测试表明,在跑步机和自行车测力计的所有运动阶段,Booster和Cortex Metalyzer 3B在VO和VE测量上没有显著差异,表明相关性良好。然而,在休息和最大运动阶段观察到Booster和Cortex Metalyzer 3B之间的测量存在差异。Booster在大多数跑步机运动阶段表现出值得称赞的可靠性和稳定性,与Cortex Metalyzer 3B系统相比,其有效性总体上可以接受。尽管如此,在休息和最大运动条件下可能会出现潜在的测量差异。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/333e/11751217/37abad067d15/fphys-15-1453942-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验