通过频域和扫频光学相干断层扫描测量青光眼患者和疑似青光眼患者的中央角膜厚度,并与标准超声方法进行比较。

Central Corneal Thickness Measurement in Patients with Glaucoma and Glaucoma Suspects by Spectral-Domain and Swept-Source Optical Coherence Tomography Compared with the Standard Ultrasonic Method.

作者信息

Machado Luz Pedro, Lubisco Valentina Mostardeiro, Leivas Lindenmeyer Rodrigo, Swarovsky Ariadne Peres, Back da Silva Anais, D Azevedo Silveira Victoria, Silva Monica Oliveira, Schopf Carolina, Lavinsky Jaco, Lavinsky Daniel, Pakter Helena Messinger, Lavinsky Fabio

机构信息

School of Medicine, Universidade do Vale do Rio dos Sinos, São Leopoldo, Brazil.

Department of Ophthalmology, Hospital de Clínicas de Porto Alegre, Porto Alegre, Brazil.

出版信息

Clin Ophthalmol. 2025 Mar 10;19:819-825. doi: 10.2147/OPTH.S512052. eCollection 2025.

Abstract

PURPOSE

To evaluate the agreement of central corneal thickness (CCT) measurements between spectral-domain (SD) and swept-source (SS) optical coherence tomography (OCT) compared with the gold standard ultrasonic corneal pachymetry (US CP).

METHODS

Adults with glaucoma and glaucoma suspects presenting with typical optic nerve head findings and high intraocular pressure with or without visual field damage were included. Patients underwent anterior segment (AS) SD-OCT (Spectralis, Heidelberg Engineering), AS SS-OCT (DRI-Triton, Topcon), and US CP. For both AS SD-OCT and AS SS-OCT, an 8.3 and a 3.0 mm single horizontal B-scan at the central cornea were performed and manually measured respectively. CCT was measured by three blinded different trained examiners. Intraclass correlation coefficient (ICC) and Bland-Altman plots were used for statistical analysis.

RESULTS

Twenty-six eyes from 16 patients with glaucoma and glacuoma suspects were included. The mean (SD) age was 59.1 (10.5) years, and 87% were women. The mean (SD) CCT measurements were 548.04 (33.11) µm for AS SD-OCT, 538.88 (34.84) µm for AS SS-OCT, and 537.12 (29.20) µm for US CP. The overall agreement ICC was 0.915 (95% CI, 0.785-0.964; p<0.001). The best agreement was found between AS SS-OCT and US CP (ICC = 0.929; 95% CI, 0.849-0.967; p<0.001). When comparing AS-OCT methods with US CP, a difference of less than 10 µm was found in 53.8% of the eyes examined by SS-OCT, and in 50% of the eyes examined by SD-OCT.

CONCLUSION

OCT emerges as an effective modality for CCT measurements. This technology provides advantages such as patient comfort, B-scan trackability, and the possibility to conduct multiple glaucoma diagnostic modalities in a single device.

摘要

目的

与金标准超声角膜测厚法(US CP)相比,评估频域(SD)和扫频源(SS)光学相干断层扫描(OCT)测量中央角膜厚度(CCT)的一致性。

方法

纳入患有青光眼和疑似青光眼的成年人,这些患者有典型的视神经乳头表现且眼压高,有或无视野损害。患者接受眼前节(AS)SD-OCT(Spectralis,海德堡工程公司)、AS SS-OCT(DRI-Triton,拓普康)和US CP检查。对于AS SD-OCT和AS SS-OCT,均在中央角膜处进行8.3mm和3.0mm的单次水平B扫描,并分别进行手动测量。CCT由三名不知情的不同训练的检查者测量。组内相关系数(ICC)和Bland-Altman图用于统计分析。

结果

纳入了16例青光眼和疑似青光眼患者的26只眼。平均(标准差)年龄为59.1(10.5)岁,87%为女性。AS SD-OCT测量的平均(标准差)CCT为548.04(33.11)µm,AS SS-OCT为538.88(34.84)µm,US CP为537.12(29.20)µm。总体一致性ICC为0.915(95%CI,0.785-0.964;p<0.001)。AS SS-OCT与US CP之间的一致性最佳(ICC = 0.929;95%CI,0.849-0.967;p<0.001)。将AS-OCT方法与US CP进行比较时,在SS-OCT检查的53.8%的眼中以及SD-OCT检查的50%的眼中发现差异小于10µm。

结论

OCT已成为一种有效的CCT测量方法。该技术具有患者舒适度高、B扫描可追踪性以及在单个设备中进行多种青光眼诊断模式的可能性等优点。