L V Prasad Eye Institute, Banjara Hills, Hyderabad, India.
Cornea. 2011 Feb;30(2):121-6. doi: 10.1097/ICO.0b013e3181e16c65.
To determine (1) repeatability of central corneal thickness (CCT) measurements by spectral domain optical coherence tomography (RTVue; Optovue, Inc, Fremont, CA) and (2) agreement between CCT measurements by RTVue and those by ultrasonic pachymetry, Orbscan, and anterior segment optical coherence tomography (ASOCT).
In a prospective observational study, 2 cohorts of normal subjects were recruited. In the first cohort (51 subjects), 5 measurements of CCT were obtained by RTVue during the same visit to determine the repeatability. In the second cohort (65 subjects), CCT measurements were obtained by RTVue, ultrasonic pachymetry, Orbscan, and ASOCT during the same visit to determine the agreement among these instruments. Repeatability was assessed by intraclass correlation coefficient (ICC), within-subject standard deviation, coefficient of repeatability, and within-subject coefficient of variation. Agreement was assessed by ICC and Bland and Altman plots.
Repeatability of CCT measurements by RTVue as assessed by ICC, within-subject standard deviation, coefficient of repeatability, and within-subject coefficient of variation was 0.99 (0.99-0.99), 2.2 (1.9-2.5), 4.2 μm (3.6-4.8), and 0.4% (0.3-0.5), respectively. The average CCT by RTVue (529 μm) was comparable to that by ultrasonic pachymetry (539 μm; P = 0.15), Orbscan (536 μm; P = 0.54), and ASOCT (526 μm; P = 0.77). The 95% limits of agreement on Bland and Altman plots ranged from 20 μm (between RTVue and ASOCT) to 33 μm (between RTVue and Orbscan).
CCT measurements by RTVue have an excellent repeatability. Although CCT measurements by RTVue are comparable to that by ultrasonic pachymetry, Orbscan, and ASOCT, the difference between instruments can be significant depending on the clinical situation considered.
(1)通过谱域光相干断层扫描(RTVue; Optovue,Inc,弗里蒙特,CA)确定中央角膜厚度(CCT)测量的重复性;(2)RTVue 与超声角膜测厚仪、Orbscan 和眼前节光相干断层扫描(ASOCT)测量的 CCT 之间的一致性。
在一项前瞻性观察研究中,招募了 2 组正常受试者。在第一组(51 名受试者)中,在同一次就诊中通过 RTVue 获得 5 次 CCT 测量值,以确定重复性。在第二组(65 名受试者)中,在同一次就诊中通过 RTVue、超声角膜测厚仪、Orbscan 和 ASOCT 获得 CCT 测量值,以确定这些仪器之间的一致性。通过组内相关系数(ICC)、个体内标准差、重复性系数和个体内变异系数来评估重复性。通过 ICC 和 Bland 和 Altman 图评估一致性。
通过 ICC、个体内标准差、重复性系数和个体内变异系数评估的 RTVue 测量 CCT 的重复性分别为 0.99(0.99-0.99)、2.2(1.9-2.5)、4.2μm(3.6-4.8)和 0.4%(0.3-0.5)。RTVue 的平均 CCT(529μm)与超声角膜测厚仪(539μm;P=0.15)、Orbscan(536μm;P=0.54)和 ASOCT(526μm;P=0.77)相当。 Bland 和 Altman 图上的 95%一致性界限范围从 20μm(RTVue 与 ASOCT 之间)到 33μm(RTVue 与 Orbscan 之间)。
RTVue 的 CCT 测量重复性非常好。虽然 RTVue 测量的 CCT 与超声角膜测厚仪、Orbscan 和 ASOCT 相当,但根据所考虑的临床情况,仪器之间的差异可能很大。