Neutral carrier sodium ion-selective microelectrode for extracellular studies.

作者信息

Ammann D, Anker P

出版信息

Neurosci Lett. 1985 Jun 24;57(3):267-71. doi: 10.1016/0304-3940(85)90503-8.

Abstract

A Na+-selective microelectrode based on a synthetic neutral carrier (ETH 157) is described. The selectivities in respect to K+, Ca2+ and Mg2+ are adequate for extracellular measurements of Na+ activities. Microelectrodes with tip diameters of about 0.7 micron have an electrical resistance of 3 X 10(10) omega and a 90% response time of less than or equal to 3 s. The drift of the potential difference of the Na+-microelectrode cell assembly in aqueous extracellular electrolyte solutions is less than or equal to 0.2 m V/3 h.