Suppr超能文献

如何在Sysmex血液分析仪上处理脂血全血细胞计数样本?

How to handle lipemic CBC samples on Sysmex hematology analyzers?

作者信息

Radišić Biljak Vanja, Dolovčak Lucija, Bakarić Iva, Nikler Ana, Saračević Andrea, Grdić Rajković Marija

机构信息

Department of Medical Laboratory Diagnostics, University Hospital "Sveti Duh", Zagreb, Croatia.

Department of Sport and Exercise Medicine, Faculty of Kinesiology, University of Zagreb, Zagreb, Croatia.

出版信息

Adv Lab Med. 2025 Mar 20;6(2):166-172. doi: 10.1515/almed-2024-0206. eCollection 2025 Jun.

Abstract

OBJECTIVES

Lipemia poses a significant preanalytical problem for complete blood count (CBC) measurement due to limited and non-standardized methods for recognition and removal. We aimed to verify the optical hemoglobin (Hb-O) measurements on the Sysmex XN-1000 hematology analyzer (HA) as a possible reliable method for managing lipemic CBC samples.

METHODS

Ninety CBC samples with varying Hb concentrations were gradually spiked with a lipid emulsion. Measurements were repeated and Hb-O concentrations were recorded. Spiked CBC samples were centrifuged (400 /10 min). Plasma was carefully removed, and Hb concentration was measured. The values obtained from the lipemic samples were adjusted according to the measurements in the plasma. The removed plasma was substituted with the analyzer's diluent, and measurements were repeated. Triglyceride concentrations were measured in lipemic plasma samples.

RESULTS

Hb-O showed statistically insignificant and acceptable bias compared to the initial Hb measurement according to the strictest acceptability criteria (-0.4 %, 95 % CI: -1.2-0.3, p=0.2447). The observed bias did not correlate with the degree of lipemia (rho=-0.072, 95 % CI: -0.295 to 0.157, p=0.537). Hemoglobin measured in samples with lipemic plasma replaced by analyzer diluent exhibited minimal, albeit statistically significant, bias (-1.1 %, 95 % CI: -2.0 to (-0.1), p=0.025). The observed bias negatively correlated with the degree of lipemia (rho=-0.369, 95 % CI: -0.550 to (-0.155), p=0.001). The highest unacceptable bias was found in the recalculated hemoglobin values based on the measured plasma hemoglobin (-3.5 %, 95 % CI: -4.1 to (-2.9), p<0.0001).

CONCLUSIONS

Hb-O measurement is the most reliable measure of lipemia removal in CBC samples on the Sysmex XN-1000 HA.

摘要

目的

由于识别和去除脂血的方法有限且不规范,脂血给全血细胞计数(CBC)测量带来了重大的分析前问题。我们旨在验证Sysmex XN - 1000血液分析仪(HA)上的光学血红蛋白(Hb - O)测量作为处理脂血CBC样本的一种可能可靠的方法。

方法

用脂质乳剂逐步添加到90份具有不同血红蛋白浓度的CBC样本中。重复测量并记录Hb - O浓度。将添加脂质后的CBC样本离心(400 / 10分钟)。小心去除血浆,并测量血红蛋白浓度。根据血浆中的测量值对脂血样本获得的值进行调整。用分析仪的稀释剂替代去除的血浆,并重复测量。测量脂血血浆样本中的甘油三酯浓度。

结果

根据最严格的可接受标准,与初始血红蛋白测量相比,Hb - O显示出统计学上无显著意义且可接受的偏差(-0.4%,95%置信区间:-1.2 - 0.3,p = 0.2447)。观察到的偏差与脂血程度无关(rho = -0.072,95%置信区间:-0.295至0.157,p = 0.537)。用分析仪稀释剂替代脂血血浆的样本中测量的血红蛋白显示出最小的偏差,尽管具有统计学意义(-1.1%,95%置信区间:-2.0至(-0.1),p = 0.025)。观察到的偏差与脂血程度呈负相关(rho = -0.369,95%置信区间:-0.550至(-0.155),p = 0.001)。在根据测量的血浆血红蛋白重新计算的血红蛋白值中发现了最高的不可接受偏差(-3.5%,95%置信区间:-4.1至(-2.9),p < 0.0001)。

结论

在Sysmex XN - 1000 HA上,Hb - O测量是CBC样本中去除脂血最可靠的测量方法。

相似文献

7
Performance evaluation of new Abbott Alinity hq hematology analyzer.新 Abbott Alinity hq 血液分析仪的性能评估。
Int J Lab Hematol. 2019 Oct;41(5):657-663. doi: 10.1111/ijlh.13082. Epub 2019 Jul 4.
10
Evaluating Interference of Lipemia on Routine Clinical Biochemical Tests.评估脂血对常规临床生化检验的干扰
J Lab Physicians. 2022 Dec 5;15(2):269-275. doi: 10.1055/s-0042-1758664. eCollection 2023 Jun.

本文引用的文献

1
Addressing the Laboratorian Shortage.应对检验人员短缺问题。
J Appl Lab Med. 2024 Jan 3;9(1):172-174. doi: 10.1093/jalm/jfad080.
2
Handling of lipemic samples in the clinical laboratory.临床实验室中脂血样本的处理
Adv Lab Med. 2023 Feb 20;4(1):5-27. doi: 10.1515/almed-2023-0003. eCollection 2023 Apr.
9
Lipemia: causes, interference mechanisms, detection and management.脂血:原因、干扰机制、检测和管理。
Biochem Med (Zagreb). 2014 Feb 15;24(1):57-67. doi: 10.11613/BM.2014.008. eCollection 2014.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验