Hartley R, Smith I J, Cookman J R
J Chromatogr. 1985 Jul 12;342(1):105-17. doi: 10.1016/s0378-4347(00)84493-x.
An improved high-performance liquid chromatographic method for the simultaneous determination of caffeine and its N-demethylated metabolites in plasma is described. Excellent resolution of all components is provided by reversed-phase chromatography using a mobile phase consisting of 1% acetic acid-methanol (83:17) at a flow-rate of 2.7 ml/min, in conjunction with a Waters Assoc. Nova-Pak C18 column which was protected by a Waters Assoc. Guard-Pak precolumn module containing a Guard-Pak CN cartridge. Rapid extraction of caffeine and the dimethylxanthines from plasma was achieved using reversed-phase octadecylsilane bonded-silica columns (Bond-Elut C18). With only 100 microliters of sample, plasma levels in the region of 50 ng/ml for the dimethylxanthines and 100 ng/ml for caffeine can be determined using ultraviolet detection at 273 nm. The method has been used for measuring umbilical cord plasma samples to provide information regarding foetal exposure to caffeine and its metabolites and is also suitable for therapeutic drug monitoring of caffeine and theophylline levels in the treatment of neonatal apnoea.
本文描述了一种改进的高效液相色谱法,用于同时测定血浆中咖啡因及其N-去甲基代谢物。使用由1%乙酸-甲醇(83:17)组成的流动相,流速为2.7 ml/min,结合沃特世公司的Nova-Pak C18柱,并由沃特世公司的Guard-Pak预柱模块保护,该模块包含一个Guard-Pak CN柱芯,通过反相色谱法可实现所有组分的出色分离。使用反相十八烷基硅烷键合硅胶柱(Bond-Elut C18)可从血浆中快速提取咖啡因和二甲基黄嘌呤。仅需100微升样品,通过在273 nm处进行紫外检测,即可测定二甲基黄嘌呤血浆水平在50 ng/ml左右以及咖啡因血浆水平在100 ng/ml左右的情况。该方法已用于测量脐带血浆样本,以提供有关胎儿接触咖啡因及其代谢物的信息,也适用于新生儿呼吸暂停治疗中咖啡因和茶碱水平的治疗药物监测。