INTRODUCTION

Modern endourology relies on Ho: YAG lasers for definitive management of urolithiasis. Current safety data is drawn predominantly from Ho: YAG laser testing in animal tissue. We aim to evaluate and characterise the grades of iatrogenic human ureteral injury due to contact with a Ho: YAG laser at commonly used lithotripsy settings.

METHODOLOGY

Human ureters were collected from consenting adults at the time of radical nephrectomy. A custom rig was developed to secure a 273micron (µm) laser fibre in contact with the urothelium for periods of 5, 10 and 20 s. Testing settings used included 8 Hz (Hz) and 800millijoules (mJ), 10 Hz/1000 mJ, and 5 Hz/1500 mJ. Grades of ureteric injury were then characterised with histopathology.

RESULTS

A total of 31 specimens were tested and the mean specimen thickness was 2.2 mm (range 0.9-3.1 mm). The depth of laser-related injury ranged from urothelium alone to full thickness perforation at all settings depending on the duration of laser contact with tissue and specimen thickness. No full thickness perforation was seen in specimens ≥ 3 mm, even at 20 s of contact duration at 10 Hz/1000 mJ. However, in specimens < 1 mm the level of injury depth ranged from urothelium to full thickness perforation at 5 s of testing.

DISCUSSION

Ho: YAG laser causes variable levels of injury, with both superficial and extensive injury noted. The thickness of the ureter may determine the extent of trauma following laser contact. Concerningly, full thickness ureter perforation may occur with commonly used laser settings at short laser contact durations.

摘要

引言

现代腔内泌尿外科依赖钬激光来明确治疗尿路结石。目前的安全性数据主要来自钬激光在动物组织中的测试。我们旨在评估和描述在常用的碎石设置下,因接触钬激光而导致的医源性人类输尿管损伤的等级。

方法

在根治性肾切除术时,从同意参与的成年人身上采集人类输尿管。开发了一种定制装置,将一根273微米(μm)的激光光纤固定在与尿路上皮接触的位置,持续5、10和20秒。使用的测试设置包括8赫兹(Hz)和800毫焦耳(mJ)、10Hz/1000mJ以及5Hz/1500mJ。然后通过组织病理学对输尿管损伤等级进行描述。

结果

共测试了31个标本,标本平均厚度为2.2毫米(范围为0.9 - 3.1毫米)。根据激光与组织接触的持续时间和标本厚度,在所有设置下,激光相关损伤的深度从仅累及尿路上皮到全层穿孔不等。在厚度≥3毫米的标本中,即使在10Hz/1000mJ的接触持续时间为20秒时,也未观察到全层穿孔。然而,在厚度<1毫米的标本中,在测试5秒时损伤深度从尿路上皮到全层穿孔不等。

讨论

钬激光会造成不同程度的损伤,既有浅表损伤也有广泛损伤。输尿管的厚度可能决定激光接触后创伤的程度。令人担忧的是,在常用的激光设置下,短时间的激光接触可能会导致输尿管全层穿孔。