Engler M B, Engler M M
West J Med. 1985 Sep;143(3):329-32.
We evaluated the effects of magnetic resonance imaging on the accuracy of three types of intravenous infusion pump devices, IVAC 530, IMED 927 and IMED 960/965. The devices were exposed to a 2,800-gauss static magnetic field at a pulsed radio frequency of 11.9 MHz operating at a maximum power of 1,200 W. Each device was tested at low, medium and fast flow rates in a controlled environment and during magnetic resonance imaging. Intravenous therapy could be carried out normally during magnetic resonance diagnostic imaging only with the IVAC 530 infusion device.
我们评估了磁共振成像对三种静脉输液泵设备(IVAC 530、IMED 927和IMED 960/965)准确性的影响。这些设备暴露于2800高斯的静磁场中,脉冲射频为11.9兆赫,最大功率为1200瓦。每个设备在受控环境中和磁共振成像期间,分别在低、中、高流速下进行测试。只有使用IVAC 530输液设备时,才能在磁共振诊断成像期间正常进行静脉治疗。