磁共振成像对静脉输液装置的影响。

The effects of magnetic resonance imaging on intravenous infusion devices.

作者信息

Engler M B, Engler M M

出版信息

West J Med. 1985 Sep;143(3):329-32.

Abstract

We evaluated the effects of magnetic resonance imaging on the accuracy of three types of intravenous infusion pump devices, IVAC 530, IMED 927 and IMED 960/965. The devices were exposed to a 2,800-gauss static magnetic field at a pulsed radio frequency of 11.9 MHz operating at a maximum power of 1,200 W. Each device was tested at low, medium and fast flow rates in a controlled environment and during magnetic resonance imaging. Intravenous therapy could be carried out normally during magnetic resonance diagnostic imaging only with the IVAC 530 infusion device.

摘要

我们评估了磁共振成像对三种静脉输液泵设备(IVAC 530、IMED 927和IMED 960/965)准确性的影响。这些设备暴露于2800高斯的静磁场中,脉冲射频为11.9兆赫,最大功率为1200瓦。每个设备在受控环境中和磁共振成像期间,分别在低、中、高流速下进行测试。只有使用IVAC 530输液设备时,才能在磁共振诊断成像期间正常进行静脉治疗。