Nair Arya Sree R, Velusamy Anand, Anand Aishwarya, Janarthanan Sridurga, Ramesh Harisharan, Chakaravarthi Thilak
MCV ENT Trust Hospital, Palghat Road, Near NGM Collage, Pollachi, Tamil Nadu 642001 India.
Caldor Health Technologies, Coimbatore, India.
Indian J Otolaryngol Head Neck Surg. 2025 Jul;77(7):2454-2460. doi: 10.1007/s12070-025-05498-1. Epub 2025 May 16.
To assess the effectiveness of a newly designed, non-contact facial nerve monitor as a cost-efficient alternative to conventional nerve monitors with facial electrodes, particularly under local anesthesia, without compromising patient safety or outcomes. Pilot study involving 49 patients aged 3-59 years, who underwent otologic surgery from December 2023 to September 2024. Five patients underwent surgery monitored by both the non-contact device and the Medtronic NIM-2, allowing comparative analysis. After calibration to industry standards, the non-contact monitor was tested on 44 additional patients, and used alone to identify the facial nerve's location and bone depth. In this study, all enrolled patients were confirmed to have no pre- operative facial palsy/paresis, establishing a uniform baseline for facial nerve function. The noncontact facial nerve monitor successfully identified the vertical segment of the facial nerve in all patients, demonstrating its reliability in nerve localization. Additionally, the noncontact monitor consistently detected the second genu, even in cases without nerve dehiscence, and accurately traced the entire horizontal segment across all patients. Adjustments in current stimulation with the monitor provided comparable accuracy to existing monitors, allowing for an approximation of bone thickness around the nerve. These results confirm our monitor's effectiveness and compatibility in identifying key facial nerve segments, offering a cost-effective and affordable alternative. Our non-contact facial nerve monitor provides a practical, affordable alternative to traditional systems, effectively identifying critical nerve segments and bone depth under a microscope. It is suitable for small private and government healthcare settings, addressing the need for accessible, low-cost surgical tool.
评估一种新设计的非接触式面神经监测仪作为传统带面部电极的神经监测仪的经济高效替代方案的有效性,特别是在局部麻醉下,同时不影响患者安全或手术效果。对49例年龄在3至59岁之间、于2023年12月至2024年9月接受耳科手术的患者进行了初步研究。5例患者在手术中同时接受了非接触式设备和美敦力NIM-2监测,以便进行对比分析。在根据行业标准进行校准后,对另外44例患者使用非接触式监测仪进行测试,并单独使用该监测仪来确定面神经的位置和骨质深度。在本研究中,所有入组患者均被确认术前无面瘫/轻瘫,从而为面神经功能建立了统一的基线。非接触式面神经监测仪成功地在所有患者中识别出了面神经的垂直段,证明了其在神经定位方面的可靠性。此外,即使在没有神经裂孔的情况下,非接触式监测仪也能持续检测到第二膝状神经节,并准确地追踪到所有患者的整个水平段。通过监测仪对电流刺激进行调整,其准确性与现有监测仪相当,能够估算神经周围的骨质厚度。这些结果证实了我们的监测仪在识别面神经关键节段方面的有效性和兼容性,提供了一种经济高效且价格合理的替代方案。我们的非接触式面神经监测仪为传统系统提供了一种实用、经济的替代方案,能够在显微镜下有效地识别关键神经节段和骨质深度。它适用于小型私立和政府医疗保健机构,满足了对可及、低成本手术工具的需求。