Boom W H, Piper D J, Ruoff K L, Ferraro M J
J Clin Microbiol. 1985 Nov;22(5):856-7. doi: 10.1128/jcm.22.5.856-857.1985.
The highly specific and sensitive latex agglutination test for cryptococcal antigen detection in cerebrospinal fluid is routine in many hospitals. Contamination of cerebrospinal fluid by a minute amount of syneresis fluid (surface condensation) from agar gave a strongly positive reaction which was heat stable, was not eliminated by pronase treatment, and was not detected by the normal rabbit globulin controls. These observations were valid for three commercially available test kits and could represent a preventable cause of some unexplained false-positive tests despite the use of adequate controls.
用于检测脑脊液中隐球菌抗原的高特异性和灵敏的乳胶凝集试验在许多医院都很常规。脑脊液被来自琼脂的微量脱水收缩液(表面冷凝物)污染会产生强烈的阳性反应,该反应热稳定,经链霉蛋白酶处理后不消除,且正常兔球蛋白对照检测不到。这些观察结果适用于三种市售检测试剂盒,尽管使用了适当的对照,但这可能是一些无法解释的假阳性检测的可预防原因。