Heelan J S, Corpus L, Kessimian N
Department of Pathology, Memorial Hospital of Rhode Island, Pawtucket 02860.
J Clin Microbiol. 1991 Jun;29(6):1260-1. doi: 10.1128/jcm.29.6.1260-1261.1991.
The latex agglutination test for Cryptococcus neoformans antigen is a simple and rapid procedure for the diagnosis of cryptococcal meningitis. Although the test is sensitive, care must be taken to prevent contamination of the sample, which may result in false-positive reactions. It was discovered in our laboratory that immersion of a platinum wire inoculating loop into a sample of cerebrospinal fluid prior to testing introduced interfering substances leading to nonspecific agglutination. After further studies, it was determined that trace amounts of surface condensation (syneresis fluid) from agar, either added to the cerebrospinal fluid or adhering to the loop, were the probable source of contamination. It is suggested that the latex agglutination test for C. neoformans antigen be performed prior to culture or on a separate sample.
新型隐球菌抗原乳胶凝集试验是诊断隐球菌性脑膜炎的一种简单快速的方法。尽管该试验很敏感,但必须注意防止样本污染,否则可能导致假阳性反应。我们实验室发现,在检测前将铂丝接种环浸入脑脊液样本中会引入干扰物质,导致非特异性凝集。经过进一步研究,确定琼脂中微量的表面缩合液(脱水收缩液),无论是添加到脑脊液中还是附着在接种环上,都可能是污染源。建议在培养前或使用单独的样本进行新型隐球菌抗原乳胶凝集试验。