Dual-isotope motion correction technique for gated exercise scintigraphy.

作者信息

Groch M W, Erwin W D, Turner D A, Domnanovich J R

出版信息

J Nucl Med. 1985 Dec;26(12):1478-84.

PMID:4067647
Abstract

In exercise multigated blood-pool imaging, significant degradation of image quality occurs as a result of patient movement under the gamma camera. Motion correction devices using centroid tracking of x-y events emanating from the organ of interest cannot be applied to blood-pool studies, because cardiac contraction and rotation masks the correctable patient motion component. We have developed a dual-isotope motion correction technique (DIMC) which utilizes a second point source of dissimilar energy (241Am) to monitor movement. Positional centroids from events incident in the 241Am window are used to develop correction coordinates which are applied to the 99mTc blood-pool events. The ability of DIMC to reduce blur due to motion has been evaluated qualitatively with phantoms and quantitatively by using spatial resolution measurements obtained from stationary line sources and from sources moving at varying rates. Based on these criteria, we have found the device to be capable of reducing over 90% of the image blur of objects moving at 5.1 cm per sec. In preliminary gated exercise studies, subjective perception of image quality was shown to be significantly improved in the DIMC corrected image, when compared to images obtained without DIMC. Improvement in image quality for exercise gated studies is of particular importance because of the low count density obtained during these procedures.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验