This report is an assessment of clinical chemistry dry-reagent methodology for veterinary use. A portable reflectance photometer and dry-reagent strips were used to measure total bilirubin, glucose, creatinine, creatine-kinase and urea in equine, bovine and canine sera. The following aspects of performance are reported: within run variation determined on animal samples, between run variation using a commercial control serum, correlation between dry-reagent and wet-reagent methodology on clinical samples. Results of the above analytes have shown a satisfactory precision and indicate an excellent correlation with the reference method of our laboratory. Minimal technical requirements, simplicity, reproducibility and accuracy comparable with wet-reagent methods are the technical advantages of this system.
本报告是对兽用临床化学干试剂方法的评估。使用便携式反射光度计和干试剂条测量马、牛和犬血清中的总胆红素、葡萄糖、肌酐、肌酸激酶和尿素。报告了以下性能方面:在动物样本上测定的批内变异、使用商业对照血清的批间变异、临床样本上干试剂方法与湿试剂方法之间的相关性。上述分析物的结果显示出令人满意的精密度,并表明与我们实验室的参考方法具有良好的相关性。该系统的技术优势在于技术要求最低、操作简单、可重复性好且准确性与湿试剂方法相当。