Buck G E, Sielaff B H, Boshard R, Matsen J M
J Clin Microbiol. 1977 Jul;6(1):46-9. doi: 10.1128/jcm.6.1.46-49.1977.
A scheme for identifying bacteria has been devised which utilizes the inhibition patterns obtained by Autobac 1 with routine and unusual antimicrobial agents and with other differentially inhibitory chemical compounds. Over 600 compounds were initially identified from the literature, and over 125 of these were selected for further testing on the basis of antibacterial activity most conducive to the instrument-generated differential scheme. Numerical growth index information derived by light scatter comparisons from the instrument were analyzed by computer, utilizing the quadratic discriminant function statistical technique. In comparison with conventional methods, accuracy for the 10 bacterial genera studied was 95% or greater. Results indicate a potential for both bacterial identification and antimicrobial agent susceptibility testing in the clinical laboratory within 3 to 5 h when using this automated approach.
已经设计出一种鉴定细菌的方案,该方案利用自动细菌鉴定仪1使用常规和非常规抗菌剂以及其他具有不同抑制作用的化合物所获得的抑制模式。最初从文献中鉴定出600多种化合物,其中125多种基于对仪器生成的鉴别方案最有利的抗菌活性被选作进一步测试。利用二次判别函数统计技术,通过计算机分析从仪器的光散射比较中得出的数值生长指数信息。与传统方法相比,所研究的10个细菌属的鉴定准确率达到95%或更高。结果表明,使用这种自动化方法在临床实验室中3至5小时内就有可能进行细菌鉴定和抗菌剂敏感性测试。