O'Brien P, Michaels H B, Aldrich J E, Andrew J W
Med Phys. 1985 Nov-Dec;12(6):799-805. doi: 10.1118/1.595788.
Clinically useful electron fields are produced on the Atomic Energy of Canada, Limited Therac 25 linear accelerator by computer-controlled scanning of the electron beam. Measurements were made to determine the properties of these electron fields. Central axis percentage depth dose and bremsstrahlung background were compared for these fields and for the fields from selected machines that use scattering foils. Dose calibrations were made in both water and polystyrene using the American Association of Physicists in Medicine Task Group 21 protocol. Measurements were made to determine the relative output factors, virtual source position, and the attenuation of the electron fields by lead.
在加拿大原子能有限公司的Therac 25直线加速器上,通过计算机控制电子束扫描产生临床上有用的电子射野。进行了测量以确定这些电子射野的特性。比较了这些射野以及使用散射箔的选定机器的射野的中心轴百分深度剂量和轫致辐射本底。使用美国医学物理学家协会任务组21协议在水和聚苯乙烯中进行了剂量校准。进行测量以确定相对输出因子、虚拟源位置以及铅对电子射野的衰减。